3-28
CL 5
Instrument Calibration
Setting Up the CL 5
Step 3:
Press to activate the control, then or
to adjust the selected place value and or to
select another v alue.
Step 4:
NOTE: Nominal and Alarm values may be overridden
in
Quality View Mode when a Custom Linear File is
active (Data Recorder Option)
(NOM. THICKNESS N/A is displayed. This value
cannot be changed.)
Press
when the adjustment is complete.
3.3 Instrument Calibration
Before measuring thickness with the CL 5, the instru-
ment and connected probe must be calibrated. It’s
important that the setup file corresponding to the
attached probe be selected prior to launching the
calibration process.
Calibration requires the use of one or more standards
of known thickness. When a contact probe is installed,
either a one or two point calibration can be specified.
Note that contact probe types require zeroing and the
zeroing process is automatically accomplished as
part of calibration. To calibrate the instrument, follow
the instructions in Figure 3-4 and on your instrument’s
display screen.
Note:
Proper calibration affects the accurate operation of
the instrument. The instrument should be calibra-
ted if there is a change in the type or temperature
of the material being tested, if the attached probe is
changed, after performing any parameter adjust-
ment, or at intervals specified in your test procedures.
3.4 Setting the Maximum and
Minimum Alarms
The CL 5 is equipped with a red LED (at the bottom
center of the instrument’s key pad) that illuminates
whenever the user-inputted minimum or maximum
Содержание Waygate Technologies CL 5
Страница 1: ...Krautkr mer CL 5 Operating Manual 021 002 296 rev 04...
Страница 2: ...CL 5 1 2 General Information 1...
Страница 10: ...CL 5 2 10 Understanding the Keypad Menu System and Displays 2...
Страница 20: ...CL 5 3 20 Setting Up the CL 5 3...
Страница 24: ...3 24 CL 5 Configuring the Instrument s Display Setting Up the CL 5 FIGURE 3 2 Selecting a Setup File...
Страница 26: ...3 26 CL 5 Configuring the Instrument s Display Setting Up the CL 5 FIGURE 3 3 Changing Parameter Settings...
Страница 35: ...CL 5 3 35 Setting Up the CL 5 Locking and Unlocking Instrument Controls THIS PAGE LEFT INTENTIONALLY BLANK...
Страница 36: ...CL 5 4 36 Measuring Thickness 4...
Страница 49: ...CL 5 5 49 Using the Optional Data Recorder 5...
Страница 56: ...5 56 CL 5 Printing a Report Using the Optional Data Recorder THIS PAGE LEFT INTENTIONALLY BLANK...
Страница 57: ...CL 5 6 57 I O Features 6...
Страница 59: ...CL 5 6 59 Transmitting Thickness Data Through an External Device I O Features...
Страница 69: ...CL 5 7 69 Specifications 7...
Страница 76: ...7 76 CL 5 Specifications CL 5 Probe Transducer Specifications THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Страница 77: ...CL 5 8 77 Maintenance 8...
Страница 79: ...CL 5 9 79 Appendix 9...
Страница 84: ...CL 5 10 84 Index 10...