background image

34

6.4  Setup-level programming

At this programming level, the 

Cfg

 menu option is now 

available. As you will recall, this option is blank at the 

operator-level (

2719

 password).

6.4.1  Configuring the display

In 

“Setting up the display” on page 30

, we discussed the 

procedures for using the disp menu to switch the LCD 

between a list of available oxygen measurement windows. 

This section provides instructions on how to create and edit 

that list. To perform this task, navigate to the 

Cfg>1:Cfg>O2

 

menu and proceed as follows:
1.  At the 

O2

 menu, press [ENTER].

2.  Use the [

] and [

] keys to highlight the 

mA

 option  

and press [ENTER].

3.  Select the outputs (0:display, 0:fault alarm, 0:analog 

output) that you wish to receive the oxygen 

measurement values in mA:
a.  Use the [

] and [

] keys to highlight a specific  

system output. (Make sure that 

0:display

 is included 

on the list).

b.  Use the [

] and [

] keys to select 

add

 or remove 

and press the [ENTER] key to change the status of the 

selected output. An output is selected if an asterisk (*) 

appears next to it.

c.  Change the status of any additional outputs that  

you wish.

d.  When the list of outputs is complete, use the 

 [

] and [

] keys to select 

OK

 and press [ENTER].

NOTE: 

You may also abort the display configuration by 

selecting cancel in step 3D above.
4.  Use the [

] and [

] keys to highlight the pcnt option  

and press [ENTER].

5.  Select the outputs (0:display, 0:fault alarm, 0:analog 

output) that you wish to receive the oxygen 

measurement values in pcnt:
a.  Use the [

] and [

] keys to highlight a specific  

system output. (Make sure that 0:display is included 

on the list).

b.  Use the [

] and [

] keys to select add or remove 

and press the [ENTER] key to change the status of the 

selected output. An output is selected if an asterisk (*) 

appears next to it.

c.  Change the status of any additional outputs that  

you wish.

d.  When the list of outputs is complete, use the [

] and 

[

] keys to select OK and press [ENTER].

NOTE:

 You may also abort the display configuration by 

selecting 

cancel

 in step 5D above.

6.5  Factory-level programming

At this programming level, all menu options are available. 

The new menus, which are available only at this level, are 

the 

cal>1:cal>adv

 menu and the opt>setup menu. Proceed 

to the appropriate section for instructions on using these 

menus.

CAUTION!

Always consult Panametrics before 

programming any of these factory-level 

features. Improper setup may impair 

the performance of the XDP system and/

or prevent it from meeting its design 

specifications.

6.5.1  The adv menu

This menu includes the following submenus:
•  Edit drift data
•  4-20mA input trim
•  AutoCal relays test
•  Zero/span cal type

Instructions for programming these features begin on the 

next page.

6.5.1a  Edit drift data

The XDP stores the results of its calibration process as drift 

data. This data is then used to compensate the live input so 

that an accurate measurement result is shown on the LCD. 

This menu provides the means to manually edit the drift data 

that resulted from the AutoCal process. To use this menu, 

navigate to the 

cal>1:cal>adv

 menu and proceed as follows:

IMPORTANT

Always consult the factory before editing the drift data via 

this menu.

1.  At the 

adv

 menu, press [ENTER].

2.  Use the [

] and [

] keys to highlight the 

edit drift data

 

option and press [ENTER].

3.  Use the [

] and [

] keys to highlight the 

zero drift

 option 

and press [ENTER].

4.  Program the following 

zero drift

 values:

a.  Use the [

] and [

] keys to highlight the 

zero drift/cal

 

option and press [ENTER]. Using the four arrow keys, 

enter the new ppm/pcnt value. Then, press [ENTER].

b.  Use the [

] and [

] keys to highlight the 

zero  

drift/total

 option and press [ENTER]. Using the four 

arrow keys, enter the new ppm/pcnt value. Then,  

press [ENTER].

Содержание Panametrics XDP

Страница 1: ...Gas analysis XDP User s manual 910 204 Rev E...

Страница 2: ...ii...

Страница 3: ...XDP Panametrics explosion proof display User s manual Attention This manual for use with standard software E 1 only 910 204 Rev E December 2014 iii...

Страница 4: ...no content intended for this page proceed to next page iv...

Страница 5: ...2 5 2 4 3 Wiring the XDP alarms J3 6 2 4 4 Wiring the calibration relays J7 6 2 4 5 Wiring the system analog output J5 6 2 4 6 Wiring the XDP analog output J8 6 2 4 7 Wiring the serial port J4 6 2 4 8...

Страница 6: ...tting up the serial port digital output 25 4 7 Using the XDP utility functions 25 4 7 1 Lock menus 25 4 7 2 Lock all 25 4 7 3 Versions 25 Chapter 5 Calibration 27 5 1 Introduction 27 5 2 Programming s...

Страница 7: ...pecifications 38 7 5 Order information 38 Appendix A CE Mark compliance 39 A1 Introduction 39 A2 EMC compliance 39 A3 LVD compliance 39 Appendix B PanaView programming 40 B 1 Introduction 40 B 2 Initi...

Страница 8: ...viii no content intended for this page...

Страница 9: ...For installations in potentially hazardous areas be sure to read the certification and safety statements document at the end of this manual before beginning the installation Auxiliary Equipment Local...

Страница 10: ...to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources we encourage you to use the appropriate take back systems Those systems will reu...

Страница 11: ...nalog input 4 20 mA 0 20 mA 0 2 V analog output RS 232 RS 485 serial port High resolution backlit LCD digital display Six key magnetic through the glass keypad Five general purpose alarm relays Automa...

Страница 12: ...no content intended for this page proceed to next page 2...

Страница 13: ...ocated within 1 8 m 6 ft of the unit Be sure to observe all installation limits and precautions described in this chapter Pay particular attention to the ambient temperature range of 10 to 60 C 14 to...

Страница 14: ...XDP system This section explains how to mount the XDP system components and the transmitter sample system at the installation site WARNING The XDP requires an external power disconnect device such as...

Страница 15: ...e slots provided to unscrew the cover Proceed to the appropriate section to wire each of the terminal blocks NOTE A long pin like plastic wiring tool is stored inside both ports of the XDP enclosure I...

Страница 16: ...used for non fail safe operation b Connect pin 5 C to the alarm device return c Connect pin 6 NO to the alarm device input 3 Connect the AutoCal indication alarm relay 3 as follows a Pin 7 NC is not...

Страница 17: ...ed off at the external disconnect device 1 Attach a cable or conduit with the three line power conductors to a cable entry port on the XDP enclosure CAUTION Be sure that the input voltage complies wit...

Страница 18: ...ATE FOUR DRIVE PINS THROUGH LABEL AND MOUNTING PLATE SECURE MARKING TO INSTRUMENT HOUSING 8 THE XDP IS AN ACCESSORY USED TO OPERATE OTHER PANAMETRICS PRODUCTS THE XDP DOES NOT DISSIPATE 40 WATTS 9 NBR...

Страница 19: ...FILE 442 1603 ON SHEET 6 REFER TO LABEL FILE 442 1604 ON SHEET 6 DETAIL E SCALE 2 1 DETAIL F SCALE 2 1 GROUND STUD PART 213 256 03 PIN SPACING 5 5 08 AWG 28 12 300V 15A CHANGES TO THIS DRAWING REQUIRE...

Страница 20: ...ZATION CONTACT THE CERTIFICATION ENGINEER 8 A 7 6 B 5 4 3 2 1 A B C 8 D 7 6 C 5 4 3 D 2 1 752 175 01 VERSION REV C D SIZE DRAWING NUMBER DRAWING NOT TO SCALE SHEET 3 OF 6 GENERATED USING SOLIDWORKS CO...

Страница 21: ...WING REQUIRE PRIOR AGENCY AUTHORIZATION CONTACT THE CERTIFICATION ENGINEER 8 A 7 6 B 5 4 3 2 1 A B C 8 D 7 6 C 5 4 3 D 2 1 752 175 01 VERSION REV C D SIZE DRAWING NUMBER DRAWING NOT TO SCALE SHEET 4 O...

Страница 22: ...TTED AT LEAST ONE FULL THREAD SHALL REMAIN FREE AT THE BASE OF THE HOLE 6X 3 4 NPT ANSI ASME B1 20 1 SEE CSS GENERAL NOTES 6X MINIMUM OF 5 THREADS REQUIRED MUST GAUGE FLUSH TO 2 TURNS LARGE USING AN L...

Страница 23: ...ive gas dust atmosphere is present 3 4 NPT conduit openings CL I DIV 1 GP B C D CL II DIV 1 GP E F G CL III TYPE 4X XDP Caution In ambients over 55 C use field wire rated for 90 C do not open when an...

Страница 24: ...S AREA TO BE RED PMS 485 C 0 M 95 Y 100 K 0 NOTES 1 MATERIAL 007 THK REF MYLAR BASE WITH REVERSE SCREENING AND 3M NO 468 PERMANENT ADHESIVE BACKING LABELL DIA 3 75 IN 2 FINISH MATTE VELVET TEXTURE 3 C...

Страница 25: ...of some common problems Compliance with the following points will help to eliminate such problems The transmitter must be properly connected to the XDP before applying power to the system Connecting o...

Страница 26: ...ly upon power up the digital display shows the following sequence of informational screens as the meter performs its internal systems checks This is a typical example of the first initialization scree...

Страница 27: ...remain unlocked when you return to measurement mode With the menus unlocked you may reenter the user program by simply pressing the ESC key only To re lock the menus see the instructions in Chapter 4...

Страница 28: ...no content intended for this page proceed to next page 18...

Страница 29: ...ting up the display Setting up the analog input Setting up the alarms Setting up the analog and digital outputs Using up the XDP utilities 4 2 Programming steps To program your XDP complete the follow...

Страница 30: ...option b Press ENTER The real time clock is set in a different menu To set the date and time navigate to the cal 1 cal setup clock menu and proceed as follows 1 At the clock menu press ENTER 2 Use th...

Страница 31: ...the end of the data line use the and keys to move to the next data point number 7 Repeat steps 5 and 6 until all of the data points have been entered 8 Press ENTER 4 4 2 Setting the analog input error...

Страница 32: ...ppm pcnt value is below this setpoint the high and high high alarms will be tripped if the current ppm pcnt value is above this setpoint When an alarm trips the corresponding symbol LL L H or HH is sh...

Страница 33: ...g output range Press ENTER Having set the basic parameters for the analog output the output should now be tested and trimmed by continuing as follows 1 Use the and keys to highlight the test option an...

Страница 34: ...Force high displays a value equal to the high point of the selected analog output range e g 20 mA for a 4 20 mA range Force low displays a value equal to the low point of the selected analog output r...

Страница 35: ...ed as follows 1 At the lock menus menu option press ENTER 2 Notice that you have exited the user program and the outline of a padlock now appears in the upper right corner of the display To program th...

Страница 36: ...26 no content intended for this page proceed to next page...

Страница 37: ...e calibration procedure the following items are required The chosen calibration gas or gases A sample system capable of providing the calibration gas or gases to the transmitter at the correct flow ra...

Страница 38: ...and keys to highlight the AutoCal types menu option and press ENTER 3 Use the and keys to highlight the desired calibration method and press ENTER Proceed to the appropriate step below for the type of...

Страница 39: ...s proceed as follows 1 Enable AutoCal as described in enabling AutoCal on page 44 2 Use the and keys to highlight the Aout indication menu option and press ENTER 3 Use the and keys to highlight the de...

Страница 40: ...m informs you of the error condition To program the auto verification function proceed as follows 1 Enable AutoCal as described in Enabling AutoCal on page 44 2 Use the and keys to highlight the auto...

Страница 41: ...AutoCal verification on page 48 it must be reset manually To do this complete the following steps 1 At the AutoCal menu press ENTER 2 Use the and keys to highlight the reset verif alarm option and pre...

Страница 42: ...32 no content intended for this page proceed to next page...

Страница 43: ...gories which require passwords that are different from each other and from the operator level password noted below and in accessing the user program on page 26 All three of the programming levels are...

Страница 44: ...xt to it c Change the status of any additional outputs that you wish d When the list of outputs is complete use the and keys to select OK and press ENTER NOTE You may also abort the display configurat...

Страница 45: ...00 Then press ENTER 6 5 1c AutoCal relays test The XDP includes four alarm relays that are related to the calibration process This menu is used to manually test the relays To use the menu navigate to...

Страница 46: ...ftware from the XDP To do this navigate to the opt setup menu on the XDP and proceed as follows 1 At the setup menu press ENTER 2 Use the and keys to highlight the erase main option and press ENTER Al...

Страница 47: ...to 10 000 ppm Input power 85 to 264 VAC 47 to 63 Hz 40 W Fuse 1 25 A Analog input 4 to 20 mA Output power supply 24 VDC 2 VDC at 1 2 A Ambient temperature range 14 F to 140 F 10 C to 60 C Keypad Magn...

Страница 48: ...B C and D Flameproof Intertek ITS12ATEX17703X II 2 G Ex d IIC T6 Gb IECEx ITS 12 0058x European compliance Complies with EMC directive 2014 30 EU 2014 35 EU LVD installation category II pollution degr...

Страница 49: ...t the cable entry closest to the chassis ground Use shielded cable to connect the power to the xdp enclosure Connect the shield to the nearest chassis ground terminal Connect the power line ground wir...

Страница 50: ...data PanaView may be used to view the live measurement data being collected by the XDP This display is in addition to the standard LCD screen on the face of the instrument There are several advantage...

Страница 51: ...is just a brief overview of how to view your live XDP data with PanaView For complete instructions on using all of the available features and capabilities refer to the PanaView user s manual Panametr...

Страница 52: ...the log file may be monitored as the data is accumulated When the log file has been completed it may be named and stored on the PC The resulting log file may be opened for viewing or printing at any...

Страница 53: ...43 Appendix C Menu maps C 1 List of menu maps Operator level cal and disp menu map Figure 14 on page 44 Operator level opt menu map Figure 15 on page 45 Factory level menu map Figure 16 on page 46...

Страница 54: ...nt Aout types ENT Test Error handlers Zero trim Span trim High high alarm High alarm Low alarm Low low alarm ENT Zero tot Zero cal View drift data Disable Enable Enable disable ENT No ENT Yes ENT Clea...

Страница 55: ...NT ENT On ENT Data length Baud rate Node ID ENT Stop bits Parity Comm type ENT see page C 1 ENT ENT User Versions Software versions X XX pcnt X XX pcnt Timed Off Span trim Zero trim Test Span setpoint...

Страница 56: ...mA input trim ENT ENT ENT Span trim Zero trim Auto verif alarm AutoCal alarm Process cal relay Trip Zero span cal relay AutoCal relays test Reset ENT Span drift Zero drift Z S setpoint Z S drift total...

Страница 57: ...4 Slot analog output Type 4 20mA Zero setpoint 0 0 Span setpoint 100 All error handlers Disabled Error handler No effect Error handler force to value 0 0 D 2 5 Alarms Trip point HH 100 H 100 L 0 LL 0...

Страница 58: ...48 no content intended for this page proceed to next page...

Страница 59: ...AutoCal setting drift limits 29 AutoCal initiating 29 AutoCal selecting type 28 B Backlight setting up 20 Basic features 1 C Cable serial port 6 Calibration 27 Calibration automatic 15 Calibration ga...

Страница 60: ...analog input error limits 21 Setting the AutoCal drift limits 29 Setting the display backlight 20 Setting up the alarms 19 22 Setting up the analog and digital outputs 19 23 Setting up the analog inpu...

Страница 61: ...ties arising from course of dealing or usage or trade Return policy If a Panametrics instrument malfunctions within the warranty period the following procedure must be completed 1 Notify Panametrics g...

Страница 62: ...ning Installation should comply with IEC EN 60079 14 Equipment is of type flameproof d design and complies with EN 60079 0 2017 EN 60079 1 2014 The product contains no exposed parts which produce surf...

Страница 63: ...EX17704X IECEx ITS 12 0062X II 2 G Ex db IIC T6 Gb IP66 22 C Tamb 65 C Caution When ambient is over 65 C use field wire rated for 90 C Do not open when explosive gas atmosphere is present NPT conduit...

Страница 64: ...bon reduction targets across critical industries including Oil Gas Energy Healthcare Water and Wastewater Chemical Processing Food Beverage and many others Join the conversation and follow us on Linke...

Отзывы: