Programming the Units

This meter comes with the option to choose 4 different
units of measure. Unless otherwise specified at the time
of the order, each meter is programmed in liters for use
with motor oil. The ‘L' will be flashing on initial start-up.
1.

Toggle the four options (‘L’, ‘QT’, ‘GAL’, ‘PT’) by

pressing the 

 button.

2.

When the desired option is on the screen, press the

 button to advance. The units of measurement

icon will stop flashing and the first digit of the scale
factor will begin flashing.

Note: If the ‘L’ units have been selected, the decimal point
will begin to flash. The user now has the option to change
the decimal point to either a period or a comma. To do

this, press the 

 button. Press the 

 button to

advance to the scale factor screen.

Changing the units of measurement from metric to English,
or from English to metric will clear the accumulated total,
and resettable total.

2. Second warning: The 

AUTO 

function will shut off and

the auto icon will disappear. This means the battery
power is too low to run the auto function. The meter
can still run in manual mode.

3. Third Warning: The screen goes blank. This means

there is no power left. The display cannot be run.
However the meter will still allow fluid to pass through
when the valve is opened, but it will not measure flow.

• The battery compartment is located on the underside of
the trigger guard. Unscrew the two screws located under
the guard and remove the battery cover to expose the
batteries.

• Replace the old batteries. This meter takes 4 AA
alkaline batteries. Replace the cover and the screws
when finished. Note battery polarity markings inside
battery compartment cover.

• Dispose of used batteries properly according to local
regulations.

Changing Factory Settings

Factory Settings

Each meter is preprogrammed and calibrated at the
factory.  Unless otherwise specified at the time of the
order, each meter is programmed in liters for use with
motor oil.

1.

Press 

 to wake up the meter if screen is blank.

To enter the programming mode, press and hold the

<PROGRAMMING>

 key located in the access hole

under the meter for 2 seconds. (See picture below)

Location of access hole for PROGRAMMING Key.

After the screen flashes, it will display the scale factor and
units of measurement.

QT

.

QT

.

Initial Programming Screen

Use a 5/32"
Allen wrench
or similar
blunt tool

Содержание RF-OMS

Страница 1: ...T This manual contains important warnings and information READ AND KEEP FOR REFERENCE Specific keypad software set up instructions are contained in the keypad packaging Model RF OMS Radio Frequency Oi...

Страница 2: ...n impaired operation or damage to equipment Radio Frequency Oil Management System Overview Badger Meter s Radio Frequency Oil Management System RF OMS consists of an A C powered keypad with an integra...

Страница 3: ...ing the system prior to installing the meter 2 Close fluid dispense valves at every dispense position 3 Once the main fluid outlet valve at the pump is closed the air pressure to the pump motor is pro...

Страница 4: ...ard Auto Series EPM QT QT Manual Mode 5 Slowly open the dispense valve and allow enough oil to pass through to ensure that the system is clean Close the valve and repeat for all dispense positions Not...

Страница 5: ...ished to reset the meter It is now ready for the next batch Do NOT press before topping off The meter will begin a new batch Normal Operating Mode Functions Total This option allows users to see the a...

Страница 6: ...n manual mode 3 Third Warning The screen goes blank This means there is no power left The display cannot be run However the meter will still allow fluid to pass through when the valve is opened but it...

Страница 7: ...slip past the gears without being detected Changing the scale factor can adjust the meter to compensate for that loss The meter multiplies each pulse by this number to correct the accuracy when it co...

Страница 8: ...perating Pressure Minimum 5 psi 35 bar Operating Temperature Maximum 120 F 50 C Operating Temperature Minimum 20 F 5 C Accuracy 0 5 0 5 5 Digit LCD Display 10 mm High x 5 mm Wide Quarts Pints Gallons...

Страница 9: ...PART DESCRIPTION PART NUMBER 1 Top Case with Screws 64103 006 2 Keypad 63802 002 1 2 NOTE For service required on parts not listed above please contact Badger Meter s authorized service center in Mil...

Страница 10: ...DESCRIPTION PART NUMBER 1 Battery Door with Screws 64103 004 2 Bottom Case with Screws 64103 003 NOTE For service required on parts not listed above please contact Badger Meter s authorized service ce...

Страница 11: ...NUMBER 1 Valve Assembly 64103 001 2 Gear Service Kit with O Ring 62896 001 3 Trigger Assembly 64103 005 1 2 3 NOTE For service required on parts not listed above please contact Badger Meter s authoriz...

Страница 12: ...ions without notice except to the extent an outstanding bid obligation exists BadgerMeter Inc P O Box 245036 Milwaukee WI 53224 9536 Telephone 414 355 0400 877 243 1010 Fax 414 355 7499 866 613 9305 w...

Отзывы: