Installation and operation manual
•
Version RCM-10/03-e
Programming Page
14/18
7.3.3 User selected full scale flow rate
This menu selection changes the full scale flow. Use
the up arrow to move the > prompt to the change
location on the screen, then press the right arrow key
to select the next menu.
Use the up or right arrow to adjust the full scale flow.
To save the value, press the Enter key. This action
will automatically move to the next menu.
This setting dictates the analog output full scale, but
does not affect the pulse per unit output.
7.3.4 Number of pulses per unit volume
This menu selection scales the pulse output and the
indicated volume total. Use the up arrow to move the
> prompt to the change location on the screen, then
press the right arrow key to select the next menu.
Use the up or right arrow to adjust the number of
pulses per unit volume. The units have been
automatically selected in the previous units menu. To
save the value, press the Enter key. This action will
automatically move to the next menu.
This setting dictates the analog output full scale, but
does not affect the pulse per unit output.
The selection range for the number of pulses for
gallons and liters is 1-1000, 1 for CCM.
7.3.5 Low flow cutoff programming
This menu selection programs when a zero condition
exists. Use the up arrow to move the > prompt to the
change location on the screen, then press the right
arrow key to select the next menu.
Use the up or right arrow to adjust the low flow cutoff.
To save the value, press the Enter key. This action
will automatically move to the next menu.