Programming Page
11/18
7.2.1 4-20 mA span adj
At this point you can adjust the 4-20 mA output span.
Use a calibrated 41/2 digit DMM set for 199.99 mA
range and attach the test leads to TB2 pins 19 and
21. Use the up and right arrow keys to change the
span value on line 2. Monitor the current indication
until the current is set for 20.00 +-.001. Then press
the enter key to store the span value and proceed to
the next menu.
7.2.2 4.20 mA zero adj
The 4-20 mA output zero can now be adjusted. Use a
calibrated 41/2 digit DMM set for 19.999 mA range
and attach the test leads to TB2 pins 19 and 21. Use
the up and right arrow keys to change the zero value
on line 2. Monitor the current indication until the
current is set for 4.00 +-.001. Then press the enter
key to store the zero value and proceed to the next
menu.
7.2.3 Programming set point
The next menu selection is the set point adjustment.
Use the up arrow to move the > prompt to the
change location on the screen, then press the right
arrow key to select the next menu.
Use the up or right arrow to adjust the set point value
from a value ranging from 0 to 100%. To save the
value, press the Enter key. This action will
automatically move to the next menu.
7.2.4 Resetting
totalizer
The next menu selection is the totalizer reset. Use
the up arrow to move the > prompt to the change
location on the screen, then press the right arrow key
to select the next menu.
Use the up arrow to clear the totalizer to zero. To
move on to the next menu press the Enter key