Meter with Remote Mount Primo

Fluid  temperature: PTFE/PFA, -40 to 302°F (-40 up to +150°C)

Chemical Injection Applications

For water line applications with a chemical injection point, the meter
should be installed upstream of the chemical injection point to
eliminate any issues with the meter performance. If an upstream
location is not possible for the meter, consider moving the injection
point downstream of the meter location. If the meter must be installed
downstream of the chemical injection point, the distance between
these (2) locations must be significant; frequently 50 - 100 feet.

When the solution made up of the water and the injected chemical
reach the Mag meter, it must be a complete homogeneous mixture.
If too close, the Mag meter will sense (2) different liquids (conductivity
different for each) and will be confused as to how to process the
information. Many other factors such as the type of injection method,
(spaced bursts versus continuous stream of drops), or whether the
chemical is injected in a liquid or gas form, can cause a wide variation
in the distance required from one application to another.

Due to this wide range of factors, it is difficult to specify a distance
that will always work for any application without establishing a value
far in excess of what would usually be required. Please contact
Badger Meter's Technical Support at 1-800-616-3837 with your
application detail and they can assist in determining if the Mag meter
should properly perform. As noted, the meter may require 50 - 100
feet from the injection point to assure the complete homogeneous
state.

Submersible Option

If the meter is to be installed in a meter vault, it should be ordered with
the remote amplifier option to be sure the amplifier is not installed
inside the vault. It is also recommended that the remote style meter
package be ordered with the submersible option (Nema 6P). This will
eliminate any potential problems from occurring should the vault
become flooded or from any presence of humidity in the vault.

Other Considerations

Avoid all pipe locations where the flow is pulsating, such as in the outlet
side of piston or diaphragm pumps.

Avoid locations near equipment producing electrical interference
such as electric motors, transformers, variable frequency, etc.

Install the meter with enough room for future access for maintenance
purposes.

The mag meter isolating liner is not intended to be used as gasket
material. Standard gaskets (not provided) should be installed to ensure
a proper hydraulic seal. When installing the gaskets, make sure they are
properly centered to avoid flow restriction or turbulence. Do not use
graphite or any electrically conductive sealing compound to hold the
gaskets in place during installation. This could affect the reading
accuracy of the measuring signal.

STRAIGHT PIPE REQUIREMENTS

For optimum accuracy performance, it is required to provide suffi-
cient inlet and outlet straight pipe runs. An equivalent to 3 diameters
of straight pipe is required on the inlet side, and 2 diameters on the
outlet side. There are no special requirements for standard concen-
tric pipe reducers.

FORWARD FLOW

MINIMUM STRAIGHT PIPE

MINIMUM STRAIGHT PIPE

ELBOW

TEE

GATE VALVE

(FULLY OPEN)

MINIMUM STRAIGHT PIPE

CHECK VALVE

GLOBE VALVE

BUTTERFLY VALVE

PUMP

ELBOW
TEE
ANY VALVE

MAGNETOFLOW MAG FLOWMETER 

(Shown with Junction Box for Remote Amplifier )

3 x D

7 x D

2 x D

D (Pipe Size)

D (Pipe Size)

STANDARD CONCENTRIC 
REDUCERS

(NO DISTANCE REQUIRED)

By using pipe reducers, a smaller meter size can be mounted in larger
pipeline sizes. This may also increase low flow accuracy performace.

In order to minimize flow disturbances and excessive loss of head,
custom fabricated pipe reducers must have a minimum slope angle
of 15 degrees. If this is not possible, install the custom pipe reducers
as if they were fittings, leaving the minimum straight pipe required.

Mounting Location

The detector should never be installed on the suction side of a pump
where a vaccuum exists to eliminate the possibility of damage to the liner.

If vibration exits on the pipeline, secure the piping before and after
the meter. If a strong vibration exists, the amplifier should be mounted
remotely.

Remote Version

Do not install cables near power cables, electrical machines, etc.
Be sure both ends of signal cable are securely fastened. Run power
and signal cables in separate conduit.

METER ORIENTATION

Magnetoflow

®

 Mag meters can operate accurately in any pipe

orientation and can measure the volumetric flow in the forward and
reverse directions (note FORWARD FLOW direction printed on the
detector label). The most recommended installation position is
vertical piping, with the liquid flowing upward (see figure below).

This installation practice ensures two objectives:

1. The pipe remains completely full of liquid, even at low flow, low

pressure applications with a discharge to the atmosphere.

2. Solid build-ups or sediments that could be part of the process

fluid will not deposit or accumulate on the liner and/or electrodes.

In horizontal piping orientation, the detector should be mounted in a
position at which the measuring electrodes axis will remain in a
horizontal plane (3 and 9 o’clock position; see figure below)

Electrode

Plane

RIGHT

Electrode

Plane

WRONG

For very low flow rate applications, the flow rate can be increased by
reducing the meter size and therefore the measuring accuracy can
be improved.

Содержание Magnetoflow Mag Meter

Страница 1: ...BadgerMeter Inc Installation Operation Manual IOM 103 02 53400 103 3 09 Magnetoflow Mag Meter Wafer Meter with Model Primo 3 1...

Страница 2: ...ing any instruc tions that may be marked on the exterior Remove all cushioning material surrounding the meter B Carefully lift the meter from the container C Retain the shipping box and all packing ma...

Страница 3: ...estriction or turbulence Do not use graphite or any electrically conductive sealing compound to hold the gaskets in place during installation This could affect the reading accuracy of the measuring si...

Страница 4: ...e following guidelines Avoid installing the detector in the highest point of the pipe line Do not install the detector in vertical downward flow sections of the pipe ON OFF valves should ALWAYS be loc...

Страница 5: ...Box Enclosure Protection For Remote Converter Option Powder coated die cast aluminum Nema 4 Cable Entries 1 2 NPT Cord Grip Meter with Primo converter Meter with junction box for remote Primo converte...

Страница 6: ...ollector Max 5Khz Digital Outputs Voltage sourcing transistor 24VDC 100mA max 3 AC electro mechanical relays 48VAC 0 5 amp max Outputs All ouputs are short circuit safe Noise Dampening Programmable fr...

Страница 7: ...ich wire goes into which terminal The chart below may be of assis tance for reference of wire color and terminal connection 2 Run cable through conduit from Primo amplifier location retaining the wiri...

Страница 8: ...ecommended that the internal fuse protection be maintained Use a 2 amp slow blow fuse How to connect the amplifier to remote digital indicators totalizers that do not require external power supply Thi...

Страница 9: ...error messages that may occur during operation of the meter The main display screen will be in one of the following modes See Flow Direction on page 11 for additional detail Uni Directional Mode Bi D...

Страница 10: ...d analog signal outputs Adjust the desired new full scale setting based on your meter size and your applications requirements making sure that it falls within the sug gested flow range of the meter Wh...

Страница 11: ...dy hit ENTER This new screen is the hysteresis value entered This value is set at the factory at 1000 mV to allow some variation in empty pipe detection to eliminate the potential for false empty pipe...

Страница 12: ...Power Up Counter This can assist in checking possible power failures Line up the with change and hit the key This will bring up the 1st error Continue hitting the key to toggle through the last 8 err...

Страница 13: ...g output ranges 4 to 20 mA 0 to 20 mA 2 to 10 mA 0 to 10 mA Press the key at the change line the following screen will appear Press or keys until you find the type of analog output that you desire amo...

Страница 14: ...en will appear Use the and keys to change the value to the desired pulse rate This value may be set from 001 to 10 000 pulses unit of measure Maximum frequency of 10kHz can not be exceeded Based on ma...

Страница 15: ...nimum flow percentage and press ENTER the following screen will appear Use the and arrows to change the maximum flow percentage and press ENTER NOTE On the signal output the relays will remain ener gi...

Страница 16: ...t be reduced by a calibration correction When finished with detector factor parameter hit ENTER and then hit key to go to the next parameter METER PIPE SIZE Place the arrow next to change and press th...

Страница 17: ...w on the nameplate reprogram to biodirectional mode Coil or electrode cables exchanged Check cable connections for cross wiring Inaccurate measuring Parameter wrong Check the parameters transmitter de...

Страница 18: ...r Contact Technical Support Water in detector Contact Technical Support Err Output Micro processor output control defective Contact Technical Support Err unknown Program error Contact Technical Suppor...

Страница 19: ...ing 1 Pull the electrode coil and display wiring plugs Loosen the screws S1 to S5 and remove the existing board assembly 2 Put in new board assembly and reassemble screws to S1 and S5 Reinstall the 3...

Страница 20: ...ions without notice except to the extent an outstanding bid obligation exists BadgerMeter Inc P O Box 245036 Milwaukee WI 53224 9536 Telephone 414 355 0400 800 456 5023 Fax 414 355 7499 866 613 9305 w...

Отзывы: