HR-E LCD 4-20
Analog Output
•
The input pulses generated within the transmitter assembly are converted to a standard 4-20 mA control signal .
•
This signal is proportional to the flow of fluid passing through the flow meter .
•
Power for the device can be obtained from a 9…50V DC control loop .
•
The default 20mA setting of the signal is defined in the registration section .
Endpoint Reading Resolution
MPORTANT
I
The standard electronic encoder output resolution of the HR-E LCD 4-20 encoder is 9 digits. Though the encoder output is 9-digit
resolution, the reading resolution sent to the reading software is dependent on the endpoint that the encoder is connected to.
Readings reported from the endpoints are the left-most significant digits of the encoder reading.
Endpoint Technology
Reading Resolution Reported to Reading Software
ORION Cellular
9-digit reading, plus the extended message capability
ORION Migratable (ME)/Fixed Network (SE)
8-digit reading, plus the extended message capability
ORION Classic (CE)
7-digit reading
See the application brief,
HR-E LCD Encoder Test Circle Codes
, available at
, for the appropriate test circle
code/reading resolutions for the encoder with ORION endpoints . Other output options are available for certain applications .
Page 10
April 2019
LCD-UM-01482-EN-07