INSTALLATION PROCEDURE

All moving parts are already mounted on the rail. One transducer is stationery, one is movable by loosening the thumbscrew 

and sliding it along the spacing scale.
The small pipe transducers are adjustable for pipe sizes between 1/2…2 in. (15…50 mm). Do not attempt to mount the 

transducers onto a pipe that is either too large or too small for the transducer. 
On horizontal pipes, mount the transducers in an orientation such that the cable exits from the side of the pipe. Do not mount 

with the cable exiting on either the top or bottom of the pipe. See 

Figure 1 on page 1

. On vertical pipes, the orientation 

does not matter.
1.  Clean the surface of the pipe. If the pipe has external corrosion or dirt, wire brush, sand or grind the mounting location 

until it is smooth and clean. Paint and other coatings, if not flaked or bubbled, need not be removed. Plastic pipes typically 

do not require surface preparation other than soap and water cleaning.

2.  Loosen the position adjustment thumbscrew and slide the transducer along the scale to set the distance between the 

them. The scale shows the distance in inches. See 

Figure 3

.

3.  Tighten the position adjustment thumbscrew hand-tight to lock the transducer in place.

1.0

2.0

1.5

0.5

Mounting Chain

Transducer Spacing Scale

Mounting Chain

Thumbscrew

Mounting Chain

Mounting Chain

Thumbscrew

Transducer Position

Adjustment Thumbscrew

(ON OPPOSITE SIDE)

Mounting

Cleat

Mounting

Cleat

Figure 3:  Transducer top view

4.  Apply contact gel to the underside of the transducers. 

Place a single bead of couplant, approximately 1/2 inch (12 mm) thick, on the flat face of the transducers. See 

Figure 4

Generally, a silicone-based grease is used as an acoustic couplant, but any good quality grease-like substance that is rated 

to not flow at the operating temperature of the pipe is acceptable. For pipe surface temperature over 130° F (55° C), use 

high-temperature paste (P.N. D002-2011-012) or non-silicone paste (P.N. D002-2011-009).

½ in.

(12 mm)

Figure 4:  Transducer bottom view

5.  Wrap the mounting chains around the pipe and secure the chains to their respective mounting cleats. See 

Figure 5

.

Installation Procedure

Page 3 

March 2021

IND-UM-02647-EN-02