1.5 Functional Overview

Contamination of refrigerants either in storage cylinders or air conditioning systems can lead to component 
corrosion, elevated head pressures and system failures when utilized by unsuspecting technicians.  The ability 
of the technician to determine refrigerant type and purity is severely hampered by the presence of air when 
attempting to utilize temperature-pressure relations.  The development of various substitute refrigerants further 
complicates the ability of a technician to identify refrigerant purity based upon temperature-pressure relation-
ships.  The substitute refrigerant blends can also introduce a fl ammability hazard to the technician and the 
ultimate end user of the air conditioning system.

The Bacharach Pur•Chek™ Advanced Refrigerant Identifi er provides a fast, easy and accurate means to deter-
mine refrigerant purity in refrigerant storage cylinders or directly in air conditioning systems.  The instrument uti-
lizes non-dispersive infrared (NDIR) technology to determine the weight concentrations of refrigerant types R12, 
R134A, R22, as well as hydrocarbons.  In addition, the unit will indicate the presence of R410A if the refrigerant 
tested matches the internal profi le established for virgin R410A.  Refrigerant purity is displayed on the LCD 
Screen.  The user must determine acceptable levels of purity based on their recovery or use standards.

Sample gas is admitted into the instrument through the supplied sample hose and presented to the sensing 
device.  The instrument provides the user with direct percent by weight concentrations of R12, R134A, R22 and 
hydrocarbons.  If the sample is determined to be pure R134A, the instrument will provide a direct readout of the 
weight percentage of air within the sample.  Note that the instrument does not consider air to be a contaminant 
since it can be removed by most refrigerant recycling equipment.  Since air is not considered to be a contami-
nant, it is possible to read 100% R134A plus 5% air.  The instrument only considers the weights of the refriger-
ant and contaminants in the total mixture for R134A as air contamination in R12 and R22 systems causes only 
minor performance degradation.

The instrument interfaces with the user with an LCD graphic display, status indicator lamps, push button com-
munication switches and an alarm horn.  Alarm indications are provided to alert of instrument fault conditions or 
contaminated refrigerant presence.  Direct percent by weight concentrations of the sample refrigerant is pro-
vided on the display as well as user directions and prompts.  An optional factory-installed printer (Model 2100-
8002) is available to print an on-the-spot analysis report.

The Bacharach Pur•Chek™ Advanced Refrigerant Identifi er provides the refrigerant technician with excellent 
knowledge of refrigerant purity and protection against refrigerant contamination and potential fl ammability.

Содержание PUR-CHEK

Страница 1: ...Product Leadership Training Service Reliability Instruction 2100 9000 Operation and Maintenance Rev 0 June 2010...

Страница 2: ...LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLI GATION LIABILITY...

Страница 3: ...r 5 3 4 Control Panel 6 3 5 Back Panel Connections 6 3 6 Hard Shell Storage Carrying Case 7 4 Operation 8 4 1 Battery Installation 8 4 2 Turning On the Unit 9 4 3 Calibration 9 4 4 Viewing the Test Re...

Страница 4: ...Pur Chek Instruction 2100 9000 ii...

Страница 5: ...or adhered to could result in minor or moderate injury IMPORTANT The use of the word IMPORTANT in this manual calls attention to a procedure practice condition or the like which if not correctly perf...

Страница 6: ...or will not emit heavy loads of oil or liquid NEVER admit any sample into the instrument at pressures in excess of 500 psig 1 3 Features and Capabilities The Pur Chek HVAC Advanced Refrigerant Identif...

Страница 7: ...levels of purity based on their recovery or use standards Sample gas is admitted into the instrument through the supplied sample hose and presented to the sensing device The instrument provides the u...

Страница 8: ...unit The chart shown is for generic reference only Displays R410A Not suitable for R404A or R507A as they are presented as R134A Blend data shown below is for reference only at 70 F and should be veri...

Страница 9: ...e will handle all of the refrigerant transfer and will provide containment up to 300psig The outer tube will provide protection of the inner tube from abrasion nicking cutting etc The hose is provided...

Страница 10: ...e soft key label at the bottom of the graphic display Red and Green LED s at the top of the Control Panel are used for visual status indications Green LED Power On Off Soft Keys Red LED 3 5 Back Panel...

Страница 11: ...ll Storage Carrying Case The hard shell storage carrying case is custom fit to the Pur Chek It provides rugged protection for the instrument as well as convenient storage for all components The enclos...

Страница 12: ...nges per hour or locate the equipment DATE CODE WARNING TO REDUCE THE RISK OF FIRE DO NOT USE THIS EQUIPMENT IN THE VICINITY OF SPILLED LISTED 6P03 Design Pressure 300 psig Refrigerant Type R 12 R 134...

Страница 13: ...internal pump This fresh air purges any excess refrigerant from the unit and ensures accurate test results Calibration requires that the hose be disconnected from the refrigerant cylinder or air condi...

Страница 14: ...concentration is greater than 90 NON represents non condensable elements such as air or another dilut ing gas 4 5 Blend Refrigerants The Pur Chek includes the Blend Chek feature for determining the pr...

Страница 15: ...Additional printouts may be made following the same procedure To exit the test press the EXIT button Figures 14 15 16 show sample printouts for various test results Bacharach Inc Refrigerant Identifi...

Страница 16: ...he help screen Figure 17 TO SET THE ELEVATION PRESS ELEV BUNNON TO SET THE LCD CONTRAST PRESS CONT Figure 19 Figure 20 ELEV CONT DONE SET FEED DONE FEET METERS 400 122 5 2 Setting the LCD Contrast The...

Страница 17: ...the new filter into the retaining clip Inspect the sample hoses for signs of oil entrapment Replacement of the sample filter usually requires cleaning or replacement of the sample hoses 5 4 Cleaning t...

Страница 18: ...on the top of the printer press the STOP button Press the TEST button to print a test message and verify that the paper is installed properly Slide the paper through the slot in the printer door and c...

Страница 19: ...re Rechargeable Battery 2100 9000 Instruction Manual 7 Service Centers Service and replacement parts can be found by contacting any of the following Bacharach Service Centers United States Bacharach I...

Страница 20: ...Pur Chek Instruction 2100 9000 16 NOTES...

Страница 21: ...Pur Chek Instruction 2100 9000 17 NOTES...

Страница 22: ...ers 621 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 Fax 724 334 5001 Website www mybacharach com Email help mybacharach com Printed in USA Registered Trademark of...

Отзывы: