4.2  Turning On the Unit

Connect the included AC power supply to the 12 VDC power input jack on the back of the unit.  Plug in the AC 
power supply to a 110/220 240 VAC outlet.

NOTE:

 If the optional battery module is installed and charged, the AC power supply is not 

required.

Press the left, soft key, power button and the splash screen shown in Figure 1 will appear for approximately 
three seconds followed by the elevation screen shown in Figure 2.  See section 5.1 for details on setting the 
elevation.  Depressing the “DONE” button will bring the Pur•Chek™ to the Calibration screen as shown in 
Figure 3.

 

SOFTWARE VERSION

XXX XX.XXX

THE OPERATING
ELEVATION HAS
NOT BEEN SET

SELECT HELP ON
THE NEXT SCREEN

READY TO AIR CAL
1. DISCONNECT
    HOSE FROM 
    TANK

2. PRESS CAL TO
    START

DONE

OFF       HELP     CAL

4.3 Calibration

Each time the Pur•Chek™ begins a new test cycle it must fi rst self calibrate.  The calibration takes 30 seconds 
(Figure 4) and brings fresh air into the unit via an internal pump.  This fresh air purges any excess refrigerant 
from the unit and ensures accurate test results.  Calibration requires that the hose be disconnected from the 
refrigerant cylinder or air conditioning system.  During calibration, the screen shown in Figure 5 will appear 
reminding the user to change the fi lter under certain conditions.  For additional details on how and when to 
change the fi lter, refer to Section 5, Maintenance and Troubleshooting.  The calibration of the unit will expire 
after approximately fi ve minutes of inactivity.  If this occurs, the screen shown in Figure 6 will be displayed 
requiring the calibration to be initiated again.

 

CALIBRATING

THIS WILL ONLY

TAKE 30 SECONDS

CALIBRATING

NOTE

CALIBRATION TIME
HAS EXPIRED
DISCONNECT HOSE
FROM TANK AND
PRESS CAL TO
RECALIBRATE

REPLACE FILTER
WHEN WHITE
ELEMENT BEGINS
TO SHOW RED
SPOTS ON OUTSIDE
DIAMETER

       CAL

Figure 4.

Figure 5.

Figure 6.

After calibrating, the unit will display the screen shown in Figure 7.  Connect the hose to the tank and press test.  
The Pur•Chek™ will display the screen shown in Figure 8.   If you wish to change any of the factory default set-
tings, refer to section 5.

BACHARACH

Figure 1.

Figure 2.

Figure 3.

Содержание PUR-CHEK

Страница 1: ...Product Leadership Training Service Reliability Instruction 2100 9000 Operation and Maintenance Rev 0 June 2010...

Страница 2: ...LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLI GATION LIABILITY...

Страница 3: ...r 5 3 4 Control Panel 6 3 5 Back Panel Connections 6 3 6 Hard Shell Storage Carrying Case 7 4 Operation 8 4 1 Battery Installation 8 4 2 Turning On the Unit 9 4 3 Calibration 9 4 4 Viewing the Test Re...

Страница 4: ...Pur Chek Instruction 2100 9000 ii...

Страница 5: ...or adhered to could result in minor or moderate injury IMPORTANT The use of the word IMPORTANT in this manual calls attention to a procedure practice condition or the like which if not correctly perf...

Страница 6: ...or will not emit heavy loads of oil or liquid NEVER admit any sample into the instrument at pressures in excess of 500 psig 1 3 Features and Capabilities The Pur Chek HVAC Advanced Refrigerant Identif...

Страница 7: ...levels of purity based on their recovery or use standards Sample gas is admitted into the instrument through the supplied sample hose and presented to the sensing device The instrument provides the u...

Страница 8: ...unit The chart shown is for generic reference only Displays R410A Not suitable for R404A or R507A as they are presented as R134A Blend data shown below is for reference only at 70 F and should be veri...

Страница 9: ...e will handle all of the refrigerant transfer and will provide containment up to 300psig The outer tube will provide protection of the inner tube from abrasion nicking cutting etc The hose is provided...

Страница 10: ...e soft key label at the bottom of the graphic display Red and Green LED s at the top of the Control Panel are used for visual status indications Green LED Power On Off Soft Keys Red LED 3 5 Back Panel...

Страница 11: ...ll Storage Carrying Case The hard shell storage carrying case is custom fit to the Pur Chek It provides rugged protection for the instrument as well as convenient storage for all components The enclos...

Страница 12: ...nges per hour or locate the equipment DATE CODE WARNING TO REDUCE THE RISK OF FIRE DO NOT USE THIS EQUIPMENT IN THE VICINITY OF SPILLED LISTED 6P03 Design Pressure 300 psig Refrigerant Type R 12 R 134...

Страница 13: ...internal pump This fresh air purges any excess refrigerant from the unit and ensures accurate test results Calibration requires that the hose be disconnected from the refrigerant cylinder or air condi...

Страница 14: ...concentration is greater than 90 NON represents non condensable elements such as air or another dilut ing gas 4 5 Blend Refrigerants The Pur Chek includes the Blend Chek feature for determining the pr...

Страница 15: ...Additional printouts may be made following the same procedure To exit the test press the EXIT button Figures 14 15 16 show sample printouts for various test results Bacharach Inc Refrigerant Identifi...

Страница 16: ...he help screen Figure 17 TO SET THE ELEVATION PRESS ELEV BUNNON TO SET THE LCD CONTRAST PRESS CONT Figure 19 Figure 20 ELEV CONT DONE SET FEED DONE FEET METERS 400 122 5 2 Setting the LCD Contrast The...

Страница 17: ...the new filter into the retaining clip Inspect the sample hoses for signs of oil entrapment Replacement of the sample filter usually requires cleaning or replacement of the sample hoses 5 4 Cleaning t...

Страница 18: ...on the top of the printer press the STOP button Press the TEST button to print a test message and verify that the paper is installed properly Slide the paper through the slot in the printer door and c...

Страница 19: ...re Rechargeable Battery 2100 9000 Instruction Manual 7 Service Centers Service and replacement parts can be found by contacting any of the following Bacharach Service Centers United States Bacharach I...

Страница 20: ...Pur Chek Instruction 2100 9000 16 NOTES...

Страница 21: ...Pur Chek Instruction 2100 9000 17 NOTES...

Страница 22: ...ers 621 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 Fax 724 334 5001 Website www mybacharach com Email help mybacharach com Printed in USA Registered Trademark of...

Отзывы: