7.3. Emptying the Water Trap Chamber
The water trap chamber should be emptied after every test, or when the water
condensate approaches the tip of the riser tube.
1. Remove water trap chamber (also see page 53).
2. Pour out liquid condensate and then reassemble trap.
7.4. Replacing the Filter Element
Replace the filter element when it becomes visibly dirty or becomes saturated
with water.
7.4.1. Material Required
•
Filter Element
(P/N 0007-1644)
•
Small Flat Blade Screwdriver
7.4.2. Procedure
1. Remove water trap chamber (also see page 53).
2. Pry apart filter chamber using a small flat-blade screwdriver. Remove
and discard old filter.
3. Install new filter and reassemble filter chamber, making sure that
surfaces “A” and “B” contact each other.
4. Reassemble trap.
Figure 7-6. Disassembling the Water Trap/Filter Assembly
Содержание PCA3
Страница 23: ...PCA 3 Manual 0024 9473 Rev 3 23 Figure 3 2 Connecting the Probe to the Analyzer...
Страница 64: ...PCA 3 Manual 64 0024 9473 Rev 3 Figure 5 23 Printer Alignment and Sample Printout...
Страница 84: ...PCA 3 Manual 84 0024 9473 Rev 3 Figure 7 2 Inside View with Rear Case Removed...
Страница 85: ...PCA 3 Manual 0024 9473 Rev 3 85 Figure 7 3 Tubing Connections Main Figure 7 4 Tubing Connections Sensors...
Страница 86: ...PCA 3 Manual 86 0024 9473 Rev 3 Figure 7 5 Wire Routing Connectors and Solder Connections...
Страница 96: ...PCA 3 Manual 96 0024 9473 Rev 3 Figure 8 3 Replacement Parts Main PCB Assembly Replacement Part Not Available...
Страница 101: ...PCA 3 Manual 0024 9473 Rev 3 101...
Страница 102: ...PCA 3 Manual 102 0024 9473 Rev 3...
Страница 103: ...PCA 3 Manual 0024 9473 Rev 3 103...