PCA
3 Manual
76
0024-9473 Rev 3
6.8. SO
2
Sensor Calibration
This procedure spans the optional sulfur dioxide sensor to a known gas
level.
6.8.1. Material Required
•
Calibration Kit
( P/N 0024-7059)
•
Gas Cylinder
(100 ppm SO
2
in Nitrogen, P/N 0024-1158)
6.8.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 66).
2. Use the ▲▼ buttons to highlight SO
2
, and then press
ENT
to display
the CALIBRATE SO
2
screen.
NOTE:
“Measured” is the current SO
2
reading, while
“Applied” is a known SO
2
level that will be applied for
calibration purposes.
3. Attach a 100 ppm SO
2
cylinder to the regulator of the calibration fixture
shown in Figure 6-12 on page 80.
4. Use the ▲▼ and
◄►
buttons to enter an applied value that exactly
equals the concentration stamped on the SO
2
cylinder.
NOTE:
The calibration range is from 9 to 1,800
ppm. An
attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
Figure 6-9. SO
2
Sensor Calibration
5. Adjust regulator for a flow meter indication of approximately 2 SCFH.
Wait until the measured reading stabilizes (approximately 3 minutes),
and then press
ENT
to calibrate the SO
2
Measured value to that of the
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
Содержание PCA3
Страница 23: ...PCA 3 Manual 0024 9473 Rev 3 23 Figure 3 2 Connecting the Probe to the Analyzer...
Страница 64: ...PCA 3 Manual 64 0024 9473 Rev 3 Figure 5 23 Printer Alignment and Sample Printout...
Страница 84: ...PCA 3 Manual 84 0024 9473 Rev 3 Figure 7 2 Inside View with Rear Case Removed...
Страница 85: ...PCA 3 Manual 0024 9473 Rev 3 85 Figure 7 3 Tubing Connections Main Figure 7 4 Tubing Connections Sensors...
Страница 86: ...PCA 3 Manual 86 0024 9473 Rev 3 Figure 7 5 Wire Routing Connectors and Solder Connections...
Страница 96: ...PCA 3 Manual 96 0024 9473 Rev 3 Figure 8 3 Replacement Parts Main PCB Assembly Replacement Part Not Available...
Страница 101: ...PCA 3 Manual 0024 9473 Rev 3 101...
Страница 102: ...PCA 3 Manual 102 0024 9473 Rev 3...
Страница 103: ...PCA 3 Manual 0024 9473 Rev 3 103...