System Setup
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
Page
4-17
4.1.2.18
Auto Return to Run Mode Time
This setting sets the time that the Mini-ICS will stay in Front Panel Calibration mode or Front Panel
User Setup mode without any user input. The purpose of this feature is that if a user enters either
mode, and forgets to exit, the setting will automatically send the unit back to run mode. The Auto
Return time may be set anywhere from 0.0 to 999.0 seconds. This setting is accessed from the
Setup Main Menu by typing “B” on the RS-232 terminal. To navigate backwards, use the <Esc> or
“Q” key on the RS-232 terminal.
4.1.2.19
Return all Settings to Factory Delivered Settings
In case of severe corruption of calibration and setting information, this setting will allow the user to
restore the Mini-ICS a
nalyzer to the “out-of-box” setting. The user may type “Y” at the prompt to
initiate a restore, or “N” and the prompt to bypass a restore. This setting is accessed from the
Setup Main Menu by typing “F” on the RS-232 Terminal. To navigate backwards, use the <Esc> or
“Q” key on the RS-232 terminal.
4.1.3 Change factory settings via Hardware Jumpers
4.1.3.1
Analog Voltage Output
The Analog voltage output must be configured using the hardware settings. In addition, the
software settings must match the jumper settings. Software changes are made via the Front Panel
User Setup menu (section 4.1.1), or the Service Port RS-232 User Setup menu (section 4.1.2.7).
4.1.3.1.1 Remove the unit from service
Make certain that all interfacing to the model Mini-ICS is disabled at the user device. Make sure
that interrupting the alarms, outputs, etc… will not interfere with normal process monitoring or
control. Disconnect power from the model Mini-ICS unit. Disconnect the removable terminal blocks
from the rear of the model Mini-ICS chassis.
4.1.3.1.2 Change jumper settings
Turn the model Mini-ICS upside down to access the jumpers through the port provided. Identify
the appropriate jumper position. Use an insulated jumper-puller, or needle nose pliers wrapped
with electrical tape to remove and replace jumpers (Figure 12).
4.1.3.1.3 Return to service
Replace cables, and terminal blocks. Reapply power. Change Analog Voltage Output setting from
front panel or service port to match new hardware settings. Perform a calibration check. Check
function of changes to ensure the new settings are recognized by the model Mini-ICS.