Installing the Analyzer
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
Page
2-4
2.1.2 Step 2
– Install the Oxygen Sensor
The Model Mini-ICS will be supplied with the Model GP, the Model CAG-250, or the Model IT
Oxygen sensor, depending on the application. In most cases, it will also be supplied with a
Neutronics-specified Sampling system. For sensor installation and pneumatic connections for the
Neutronics Inc. Sampling system, please refer to the Sampling System Manual.
2.1.3 Step 3
– Install the Analyzer
DANGER:
Electrical connections on the rear of the Model Mini-ICS Oxygen
analyzer may have hazardous voltages present once power has been applied to the
unit. High voltages may remain present for a short time even after power has been
disconnected from the analyzer. Take care in observing standard electrical practices
when making electrical connections to the Model Mini-ICS Oxygen analyzer.
DANGER:
The model Mini-ICS analyzer is not rated intrinsically safe or explosion proof. Be
certain that no flammable gases are present in the area where the Model Mini-ICS analyzer will be
installed.
CAUTION:
The model Mini-ICS housing is not rated waterproof. Do not mount the analyzer
or the sensor in an area where it may contact water or other liquid elements
.
WARNING:
Be certain that all power is OFF to the analyzer and associated wiring (cables)
before attempting installation. DO NOT WORK WITH LIVE WIRES! Do not leave any exposed
wire at the terminal blocks. Before applying power, ensure terminal blocks are fully inserted into
the mating connector at the analyzer.
A label depicting the terminal block arrangement is affixed to the top of the chassis for easy
reference during installation and maintenance. The terminal blocks feature screwed terminals.
The terminal blocks are also removable for ease of wiring or removal of the analyzer module.
Please reference Figures 8 and 9 below, and drawing SP-E-1473, in the rear of this manual,
showing the detail of the electrical interface terminals, and wiring connections.
Figure 8
–
analyzer electrical interface
label (Vac version shown)