
Multi-Zone Gas Monitors
2
P/N: 3015-5074 Rev 11
WARNI NG:
A switch or circuit breaker must be included in the building installation. The
switch must be in close proximity to the monitor and within easy reach of the operator. The
switch must be clearly marked as the disconnecting device for the equipment.
WARNI NG:
Under no circumstances should the monitor be operated without connection
to a protective ground. Doing so poses a potential shock hazard and is also a violation of
electrical safety standards applicable to this type of equipment.
WARNI NG:
Do not operate this equipment in the presence of flammable liquids, vapors,
or aerosols. Operation of any electrical instrument in such an environment constitutes a
safety hazard.
WARNI NG:
It is imperative that the exhaust port on this instrument be properly vented as
described in this manual. Failure to do so may constitute a safety hazard.
WARNI NG:
Extreme care should be exercised when accessing the interi or of the monitor.
Only qualified electrical maintenance personnel should make connections and perform
adjustments. Always remove A C power before opening the monitor’s enclosure.
WARNI NG:
The protection provided by the monitor may be impaired if the monitor is used
in a manner not specified by Bacharach, Inc. Modifications to this monitor, not expressly
approved, will void the warranty.
WARNI NG:
Do not continue to use this equipment if there are any symptoms of
malfunction or failure. In the case of such occurrence, de-energize the power supply and
contact a qualified repair technician or the nearest Bacharach Service Center.
WARNI NG:
This devic e uses type “F” fuses (F1 and F2) rated at 1.0 A, 250 VAC.
Replace ONLY with Bac harach-approved fuses.
WARNI NG:
Electrical installation should be performed by a certified electrician, and must
comply with all applicable NEC/ CEC and local electrical safety codes.
IMPORTANT:
Use ONLY the provided knockouts for electrical and communications
wiring. Drilling into the box will void the warranty.
IMPORTANT:
This device is classified as Installation Category II, Pollution Degree II, as
defined by UL.
IMPORTANT:
This device is designed for operation at or below an altitude of 6,562 ft
(2,000 m). Do not operate this device above this altitude limit.
NOTE:
To clean the outside of the case
use a dry cloth. To avoid shock hazard and/or
equipment damage, DO NOT use soap and water.
Содержание MultiZone
Страница 52: ...Multi Zone Gas Monitors 46 P N 3015 5074 Rev 11 ...
Страница 58: ...Multi Zone Gas Monitors 52 P N 3015 5074 Rev 11 ...
Страница 60: ...Multi Zone Gas Monitors 54 P N 3015 5074 Rev 11 ...
Страница 71: ...Multi Zone Gas Monitors P N 3015 5074 Rev 11 65 APPENDIX C SYSTEM MENU MAP ...
Страница 72: ...Multi Zone Gas Monitors 66 P N 3015 5074 Rev 11 ...
Страница 73: ...Multi Zone Gas Monitors P N 3015 5074 Rev 11 67 APPENDIX D AGENCY APPROVALS ...
Страница 74: ...Multi Zone Gas Monitors 68 P N 3015 5074 Rev 11 ...
Страница 76: ...Multi Zone Gas Monitors 70 P N 3015 5074 Rev 11 ...
Страница 78: ...Multi Zone Gas Monitors 72 P N 3015 5074 Rev 11 ...