Calibration and Maintenance
0024-9486 Rev 2
55
Step
T-STACK Calibration Procedure
"Measured" is the current temperature reading. "Applied" is a known
temperature that will be applied for calibration purposes.
4
Set thermocouple simulator to 32° F (0° C), and then use the UP (
), DOWN
(
), LEFT (
), and RIGHT (
) arrow buttons to enter an Applied value that
exactly equals the setting of the simulator.
Alternatively:
Submerge probe tip into an ice-water bath with a
thermometer, wait several minutes, and then use the UP (
) and DOWN (
)
arrow buttons to enter an Applied value that exactly equals the thermometer
reading.
NOTE:
The calibration range is from 32 to 41° F (0 to 5° C). An
attempt to calibrate outside this range will cause the message
“Applied Value High" (or Low) to appear at the bottom of the
screen.
5
Wait until the Measured reading stabilizes, and then press ENTER to calibrate
the TS-Zero Measured value to that of the Applied value, after which the
message “Good Calibration” should briefly appear followed by the CALIBRATE
TS-SPAN screen.
6
Set thermocouple simulator to 572° F (300° C), and then use the UP (
),
DOWN (
), LEFT (
), and RIGHT (
) arrow buttons to enter an Applied
value that exactly equals the setting of the simulator.
Alternatively:
Submerge probe tip into a container of boiling water with a
thermometer, wait several minutes, and then use the arrow buttons to enter
an Applied value that exactly equals the thermometer reading.
NOTE:
The calibration range is from 175 to 625° F (80 to
329° C). An attempt to calibrate outside this range will cause
the message “Applied Value High” (or Low) to appear at the
bottom of the screen.
Содержание Fyrite INTECH
Страница 4: ...Fyrite InTech Manual iv 0024 9486 Rev 2...
Страница 11: ...Fyrite InTech Manual Overview 0024 9486 Rev 2 5 1 6 Components...
Страница 12: ...Overview Fyrite InTech Manual 6 0024 9486 Rev 2...
Страница 16: ...Overview Fyrite InTech Manual 10 0024 9486 Rev 2...
Страница 56: ...Calibration and Maintenance Fyrite InTech Manual 50 0024 9486 Rev 2 5 3 Water Trap and Filter Replacement...
Страница 70: ...Declaration of Conformity Fyrite InTech Manual 64 0024 9486 Rev 2...
Страница 71: ...Fyrite InTech Manual 0024 9486 Rev 2 65...