Calibration and Maintenance
Fyrite
InTech
Manual
54
0024-9486 Rev 2
NOTE:
Bacharach offers a convenient Exchange
Program (where available) that allows the
customer to regularly receive pre-calibrated
replacement sensors that include a code that can
be entered into the analyzer for a quick convenient
setup. Contact Bacharach customer service for
more details about this program.
5.5. T-Stack Calibration
This procedure first zeroes and then spans stack temperature to known
temperature values.
The use of an electronic thermocouple simulator is the preferred method of
producing the desired calibration temperatures. Alternatively, ice and boiling
water baths can be used.
5.5.1. Materials Required
•
Thermocouple simulator (K-type)
Range:
0 to 600° F (-18 to 316° F)
Accuracy:
± 0.5° F (± 0.3° C)
•
(Alternatively) ice water, boiling water, thermometer
5.5.2. T-Stack Calibration Procedure
Step
T-STACK Calibration Procedure
1
Plug the simulator into the T-STACK connector located at the bottom of the
analyzer.
Alternatively:
Plug the probe's thermocouple into the T-STACK connector
located at the bottom of the analyzer.
IMPORTANT:
DO NOT attach the probe's gas hose to the
analyzer's GAS port; otherwise water will be drawn into the
analyzer!
2
If not already done, turn ON the analyzer and display the CALIBRATION
Menu. Note that this requires password validation (see page 34).
3
Use the UP (
) and DOWN (
) arrow buttons to highlight T-Stack, and then
press ENTER to display the CALIBRATE TS-ZERO screen.
Содержание Fyrite INTECH
Страница 4: ...Fyrite InTech Manual iv 0024 9486 Rev 2...
Страница 11: ...Fyrite InTech Manual Overview 0024 9486 Rev 2 5 1 6 Components...
Страница 12: ...Overview Fyrite InTech Manual 6 0024 9486 Rev 2...
Страница 16: ...Overview Fyrite InTech Manual 10 0024 9486 Rev 2...
Страница 56: ...Calibration and Maintenance Fyrite InTech Manual 50 0024 9486 Rev 2 5 3 Water Trap and Filter Replacement...
Страница 70: ...Declaration of Conformity Fyrite InTech Manual 64 0024 9486 Rev 2...
Страница 71: ...Fyrite InTech Manual 0024 9486 Rev 2 65...