72
Instruction 3015-4275
Status Register
Register 0x0011h
R/W
10 bytes
Variable Type
Length
Description
Mode
UC
1 byte
Defines Operating Mode of AGM300. 0 = normal Mode; 1
= Zone_Hold Mode; 2 = Diagnostic Mode; 3 = Service
mode.
DO NOT MODIFY (use zone hold register or
service mode register to change this parameter)
State
UC
1 byte
Defines AGM300 Current State. 0 = Idle; 1 = Sampling; 2
= Zeroing; 3 = Warm Up, 4 = Pressure Check
DO NOT
MODIFY
Measuring
UC
1 byte
Value = 1 if unit is acquiring detector signal for running avg.
DO NOT MODIFY
Active_Zone
UC
1 byte
Current Zone being checked. 0=zone1, 1=zone2, etc.
Max _Alarm
UC
1 byte
Indicates highest non-acknowledged alarm level
DO NOT
MODIFY
Alarm_Count
UC
1 byte
Number of alarms that are currently active
DO NOT
MODIFY
UNUSED UC
1
byte
UNUSED
Loop_Card
UC
1 byte
Value = 1 if 4-20mA card has been detected
DO NOT
MODIFY
Fault
UI
2 bytes
See Note Below
Fault Flag Structure uses bitwise access to 16 bit word as defined below:
Bit 15 (MSB) Clipping Fault
A/D out of range
Bit 14
Zero Fault
Zero voltage outside factory limits
Bit 13
No Flow
No flow on any zone
Bit 12
Purge Flow
No flow on purge
Bit 11
Zone Flow
No flow on a particular zone(s)
Bit 10
Trigger Fault
IR Source clock trigger missing
Bit 9
Unused
Bit 8
Unused
Bit 7
Config. Fault
No Zones Enabled
Bit 6
Unused
Bit 5
Unused
Bit 4
Loop Fault
Open Current loop
Bit 3
RS-485 Fault
Comm Error
Bit 2
Pressure Fault Pressure out of normal operating range
Bit 1
Bench Temp Fault
Bench temperature out of normal operating range
Bit 0
(LSB) Box Temp Fault Box temperature fault
Содержание ADM800
Страница 17: ...Instruction 3015 4275 9 AGM300 AC Input Power and Ground Connections...
Страница 33: ...Instruction 3015 4275 25 ADM800 AC Input Power and Ground Connections...
Страница 70: ...62 Instruction 3015 4275 Notes...
Страница 71: ...Instruction 3015 4275 63 Appendix...
Страница 74: ...66 Instruction 3015 4275 ADM800 Logic Diagram...
Страница 93: ...Instruction 3015 4256 85 Notes...