Operation
N
2
O Portable Monitor
14
Instruction 3015-4787
3.10 Working with System Faults
3.10.1 Functional Overview
If a system malfunction occurs (see Fault Code list, page 15), the front panel
FAULT
light will flash and
the audible alarm, if activated (refer to
AUDALRM
function page 11), will start beeping.
3.10.2 Clearing / Silencing a Fault Alarm
The
FAULT
light and audible alarm will automatically turn OFF after the cause of the fault has been
eliminated.
Pressing the
SILENCE
button while a fault condition still exists causes the internal audible alarm to turn
OFF for a period of time as set by the
SILENCE
function (page 12). The front panel
FAULT
light will
continue to flash as a reminder that a fault condition still exists. At the end of the silence period if the
cause of the fault has not been corrected the audible-alarm circuit will reactivate.
3.10.3 Viewing the Faults Log
From the
Data Display
screen, press any of
the
Keypad
buttons to display the first
Function Menu
screen. Next, use the
Keypad
buttons to move the arrow (>) until it points to the
FAULTS
function, and
then press the
ENTER
button to display the
Faults
Log
screen.
The
Faults Log
screen shows the monitor’s current fault status. If the
fault is still present when the
FAULTS
function is selected, then the
current cause of the fault is displayed along with the date and time it
occurred. If the cause of the fault has been cleared, then the
Faults Log
screen will show <0000> along with the date and time the fault was
cleared. Use the
Keypad
buttons to scroll through the fault log. In the
screens shown to the left, record #03 shows that a Sample Flow Fault (fault
code <0800>) occurred on 07/11/05 at 08:17, while record #04 shows that
the fault was cleared on 07/11/05 at 9:00.
The cause of the fault is identified by a numeric fault code. To convert the
fault code into a text description of the fault, first press the
ENTER
button
and then use the
Keypad
buttons to scroll through the display until the
text description of the fault appears. If the fault code is a combination of
two or more faults, then continue to use the
Keypad
buttons until all fault
text descriptions have been displayed. For example, the fault code <1800>
represents the combination of both a Sample Flow <0800> and a Purge
Flow <1000> fault as shown to the left.
The Fault Log can be cleared as described in the “
Working with the Gas Alarm" section.
Gas-Alarm Light
If the detected gas level reaches the gas-alarm level as set by the
ALRMLVL
function, the front panel
ALARM
(red) light will flash and the audible alarm, if activated (refer to
AUDALRM
function on page
11), will start beeping. The ALARM light and audible
alarm
will automatically turn OFF once the
detected gas level drops below the gas-alarm level.
Silencing a Gas Alarm
Pressing the
SILENCE
button while the alarm circuit is activated causes the audible alarm to turn OFF
for a period of time as set by the
SILENCE
function (page 12). The front panel
ALARM
light will continue
to flash as an indication that an alarm condition still exists. At the end of the silence period if the detected
gas level is still above the gas-alarm level the audible alarm will reactivate.
FAULT CODE <1800>
12 SAMPLE FLOW
FAULT CODE <1800>
13 PURGE FLOW
#03 <0800> @
07/11/05 08:17
#04 <0000> @
07/11/05 09:00