background image

N

2

O Portable Monitor 

Operation 

Instruction 3015-4787 

 

11 

During normal operation the 

Data Display

 screen shows when the monitor is performing the following 

two functions: 

MEASURE

 indicates that the monitor is actively measuring gas. In the screen shown above the monitor is 

currently detecting 100 ppm of N

2

O gas. The 

2ndUNIT

 value (Blank, Peak PPM, Continuous TWA, or 

8 Hr TWA) is user-selected (refer to 

2ndUNIT 

function on page 13). The measurement cycle runs for 

4 minutes. 

PURGE

 is displayed when the monitor is resetting its infrared detector to a baseline of 0 ppm using the 

fresh air in the monitor’s internal purge-air bag. The purge cycle runs for 10 seconds.  
 

3.9.4  Function Screens 

The Function

 

screens are used to set up the monitor and display stored data. 

From the 

Data Display

 screen, press both

 

the

 Keypad

 

Left

 and 

Right

 buttons at the same time to 

display the first 

Function Menu

 screen. Next, use the 

Keypad

 buttons to move the arrow (>) until it 

points to the desired function, and then press the 

ENTER

 button to select that function.  

 
 
 
 
 
Once a function has been selected, use the 

Keypad

 to scroll through the displayed data or to change a 

parameter associated with that function. Press 

ENTER

 to save any newly entered parameters. Press the 

ESC 

button to return to the previous screen without saving. 

Note that if no buttons are pressed within 90 seconds after selecting a function, the unit returns to the 

Data Display

 screen. 

FILLBAG

 – Press 

ENTER

 to manually fill the purge-air bag. 

WARNING:

 

Fill the purge-air bag in a clean, fresh-air environment 

that is at the same temperature and humidity as the area being 

checked for gas. DO NOT fill the bag in an area that is 

contaminated with N

2

O gas.

 

EMTYBAG

 – Press 

ENTER

 to manually empty the purge-air bag. Use 

this function to drain the bag if it becomes contaminated with gas. 

ALRMLVL

 – Detected gas level that causes the front panel 

ALARM

 

(red) light to turn ON and the audible alarm to sound (if AUDALRM 

function is set to ALARM or ALARM/FAULT).  

 

AUDALRM

 – The internal audible alarm is programmable to function as 

follows:  

 

OFF

 – Audible alarm is disabled. 

 

ALARM

 – Alarm sounds when detected gas level exceeds the ALRMLVL level as evidenced by the front 

panel ALARM (red) light turning ON. 

>FILLBAG       EMTYBAG  

  ALRMLVL

  

AUDALRM

      

 

>PPM LOG          LOG INT 

  FAULTS

     

DIAG

 

 

>SILENCE             P-CHK   

  GAS                   CLOCK 

 

FILLING BAG...  

DRAINING BAG...  

AUDIBLE ALARM IS 

      ALARM 

EDIT ALARM LEVEL 

       0300 ppm 

>2ndUNIT        PRINT 

  PLOT

     

PC-DUMP

 

 

Содержание 3015-4790

Страница 1: ...table Monitor Portable Area Gas Monitor PAGM with Infrared Sensor for N2O Gas Instruction 3015 4787 Operation Maintenance Rev 8 February 2013 Product Leadership Training Service Reliability Patent 6 5...

Страница 2: ...N AND ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIG...

Страница 3: ...al 6 3 3 Gas Detector Operation 6 3 4 Time Weighted Average TWA Operation 8 3 4 1 Overview 8 3 4 2 Starting a New or Continuing TWA twa 8 3 5 Peak Reading Operation 9 3 6 Leak Detector Operation 9 3 7...

Страница 4: ...ta to a Personal Computer 20 3 17 Importing Saved Data Into a Spreadsheet 21 4 MAINTENANCE 22 4 1 Disassembly 22 4 2 Internal Particulate Hydrophobic Filter Replacement 23 5 PARTS AND SERVICE 24 5 1 R...

Страница 5: ...ich if not correctly performed or adhered to could result in personal injury or death 1 3 Caution Statements The use of the word CAUTION in this manual denotes a potential hazard associated with the u...

Страница 6: ...f the monitor is used in a manner not specified by these instructions Changes or modifications to this monitor will void the warranty 1 5 5 Altitude Limit 6 562 ft 2 000 m 1 5 6 Cleaning Clean the out...

Страница 7: ...Overview The Bacharach Nitrous Oxide N2O Portable Monitor is designed to detect measure and display the concentration of N2O gas in the area being monitored To ensure accurate gas measurements the mon...

Страница 8: ...to 50 C Ambient Humidity 5 to 90 RH non condensing Altitude Limit 6 562 ft 2 000 m Power DC power pack provides 8 12 hours of operation Power Consumption 15 Watts Front Panel 3 Indicator lights ON Gre...

Страница 9: ...ing onto the elbow connector and attach the elbow onto the Gas Sample Port on the front of the monitor as shown in the illustration below The gas sample line can be up to 50 ft 15 2 m in length All tu...

Страница 10: ...filter To connect the exhaust line to the exhaust port s PTC fitting firmly push the appropriate tubing into the hole in the center of the connector until it seats in the connector Refer to the figur...

Страница 11: ...ir bag The purge cycle runs for 10 seconds The monitor will operate for approximately 40 minutes before the purge air bag needs refilled When the message PURGE AIR EMPTY appears take the monitor to a...

Страница 12: ...ll continue to be updated twa lower case letters is continuously calculated over the operating time of the monitor This mode is useful for monitoring an area for a few minutes or more and is the prefe...

Страница 13: ...ndby Mode When taken out of standby the monitor is ready for operation without waiting an additional 15 minutes to warm up While in standby the monitor will stop taking gas samples and stops purging i...

Страница 14: ...firmware revision level After a brief moment the Warm Up screen is displayed along with the front panel ON light green blinking The monitor requires 15 minutes to warm up from a cold start after whic...

Страница 15: ...ugh the displayed data or to change a parameter associated with that function Press ENTER to save any newly entered parameters Press the ESC button to return to the previous screen without saving Note...

Страница 16: ...PPM Log can be cleared per Section 3 11 LOG INT Sets the interval at which measurements are logged to memory from 1 to 9999 minutes Factory default is 1 minute The logged measurements can be viewed us...

Страница 17: ...ed to appear in the upper right hand corner of the Data Display Screen NONE Blank Peak PPM Peak gas reading Cont twa Time Weighted Average calculated on a continuous basis 8hr TWA Time Weighted Averag...

Страница 18: ...rough the fault log In the screens shown to the left record 03 shows that a Sample Flow Fault fault code 0800 occurred on 07 11 05 at 08 17 while record 04 shows that the fault was cleared on 07 11 05...

Страница 19: ...minated with gas Take monitor to a clean air area and use the EMTYBAG and FILLBAG functions several times to decontaminate the purge air bag 0200 Gain Set Fault The digipot autotune sequence has faile...

Страница 20: ...level the audible alarm will reactivate 3 12 Clearing the PPM Faults Data Up to 200 gas measurements and 30 fault events are stored by the monitor To clear stored data first display the data to be cl...

Страница 21: ...isk The purge valve can be opened and closed by pressing the Keypad Left button An asterisk appears on the display when the purge valve is open causing the monitor to draw air from its purge air port...

Страница 22: ...d closed Note that an asterisk appears when the purge valve is open causing the monitor to draw air from the purge air bag The Keypad Down button toggles the pump ON and OFF Pressing the ENTER button...

Страница 23: ...s using the LOG INT function page 12 2 Interconnect the RS 232 connectors located on the printer and monitor 3 If not already done refer to the printer manual and set the printer s communication param...

Страница 24: ...to be made as follows a Click Start click Run In the Run box type hypertrm exe and click OK The New Connection dialog box should appear b Type in a New Connection Name such as N2O Click OK to select...

Страница 25: ...eadsheet A text file that was created as described in Section 3 17 can be imported into spreadsheet programs that are capable of importing comma delimited files The following describes how to generate...

Страница 26: ...or disassemble the monitor s metal chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Remove monitor and its battery pack from the soft carrying case 2 Unplug battery pack...

Страница 27: ...ement filter P N 0007 1654 Medium Phillips head screwdriver Procedure 1 Disassemble the monitor s metal chassis per Section 4 1 2 See illustration below and locate the internal particulate hydrophobic...

Страница 28: ...ft 0003 6182 Replacement Pump 3015 4239 Battery for main PCB board 0204 0020 Replacement Bench Kit 3015 5247 Optional Accessories Item Description Part Number IrDA Printer Kits Parts Accessories 120...

Страница 29: ...tacting any of the following Bacharach Service Centers United States Bacharach Inc 621 Hunt Valley Circle New Kensington PA 15068 Phone 724 334 5051 Fax 724 334 5723 Email help mybacharach com Canada...

Страница 30: ...Declaration of Conformity N2O Portable Monitor 26 Instruction 3015 4787 6 Declaration of Conformity...

Страница 31: ...N2O Portable Monitor Instruction 3015 4787 27 Notes...

Страница 32: ...unt Valley Circle New Kensington PA 15068 7074 Toll Free 800 736 4666 Tel 1 724 334 5000 FAX 1 724 334 5001 Website www MyBacharach com E mail help MyBacharach com Printed in U S A Registered Trademar...

Отзывы: