B9Creations B9Clean Скачать руководство пользователя страница 16

 

 

 

 

 

 

7.5 

Configuring the cleaning cycle settings 

 

The Start/Stop button (A) starts or stops a cleaning cycle and is 

located on the right side of the control interface. A cleaning cycle 

is characterized by the 3D printed parts in the inner container 

being submerged in isopropyl alcohol that is stirred at a specified 

speed for a specified duration.

 

 

The duration of the cleaning cycle is measured in minutes and 

controlled by the top set of buttons and indicator lights (B) on 

the left side of the control interface.

 

 

The speed at which the isopropyl alcohol is stirred is controlled 

by the bottom set of buttons and indicator lights (C) on the left 

side of the control interface.

 

 

Use the buttons on either side of the duration and speed 

indicator lights to configure the cleaning cycle settings as desired.

 

 

 

7.6 Overview of starting, stopping or adjusting the cleaning 

cycle.   

 

Before starting the cleaning cycle, ensure the following 

conditions are met.

 

 

The outer container, inner container, inner container plug, 

hanging arm, Build Table, and lid are all secured in their proper 

positions. 

 

The isopropyl alcohol fill level, inner container plug, and hanging 

arm are correct for the height of the 3D printed parts to be 

cleaned. 

 

The magnetic stirring bar is placed in the bottom of the inner 

container. 

 

The part tray is inside the inner container only if any one of the 

3D printed parts to be cleaned is not attached to the Build Table.

 

The control interface’s indicator lights show the desired settings.

 

 

Press the Start/Stop button to begin the cleaning cycle. The 

cleaning cycle will run using the duration and spin speed set on 

the control interface, reversing the direction of the spin every 

minute. Cleaning process timer count down and stirrer operation 

starts after filling of internal container. 

 

 

To alter the duration or spin speed while the cleaning cycle is in 

progress, change the desired settings on the control interface.

 

 

Содержание B9Clean

Страница 1: ...USER MANUAL B9Clean...

Страница 2: ...nd or registered trademarks of the respective owners The publisher and the author make no claim to these trademarks While every precaution has been taken in the preparation of this document the publis...

Страница 3: ...etup 12 7 Operation 13 7 3 Positioning the inner container drain plug 15 7 4 Placing the parts in the inner container 15 7 5 Configuring the cleaning cycle settings 16 7 6 Overview of starting stoppin...

Страница 4: ...Technical Specifications 20 10 2 Operating conditions 20 10 3 Information on Manufacturer and Electrical Specification Product Label 21 11 Troubleshooting Guide 21 12 Service 24 13 Quick Reference 24...

Страница 5: ...pply to Type of machine Description B9Clean Enclosed isopropyl alcohol agitator for cleansing resin based 3D printed parts of excess uncured resin The B9Clean is an easily operated laboratory grade 3D...

Страница 6: ...tage Failure to comply with safety guidelines and procedures can result in electrical shock that may lead to injury or death FLAMMABLE This symbol warns of the dangers of flammable compounds Failure t...

Страница 7: ...cumented in this book 2 5 Safety Information Safety concerning the protection of persons the environment and the material to be processed mainly depends on the behavior of the persons operating the un...

Страница 8: ...f the required tests maintenance and repair work have been performed by B9Creations LLC or trained and authorized service staff NOTE This equipment has been tested and found to comply with the limits...

Страница 9: ...Proper Operation CAUTION The unit is suitable to be set up and operated in clean laboratory environments The unit has been designed for continuous operation Only 3D printed parts made of cured resin...

Страница 10: ...on upon delivery of the unit If items are missing or damaged contact your supplier 4 1 Scope of Delivery Build Table included for B9 Core Series owners only A B9Clean Unit B Quick Start Guide C Access...

Страница 11: ...f this User Manual NOTE If requesting technical assistance always take note concerning the model number and serial number of the unit 5 2 Control Interface Base Front 1 Start Stop button 2 Cleaning du...

Страница 12: ...unter liquid or fluid leaks The unit shall be installed in a low dust environment When not in use the containers shall be placed properly in the base and the lid shall be secured The room temperature...

Страница 13: ...onnect the unit to any type of emergency backup generator that is not connected to mains power NOTE If additional hardware is installed it could cause malfunctions to the system Do not install any oth...

Страница 14: ...age the B9Clean s electronics or surrounding materials FLAMMABLE Isopropyl alcohol is highly flammable in both liquid and vapor states Read the safety information in Section 2 5 before handling isopro...

Страница 15: ...s can be cleaned while still attached to the Build Table or while resting in the part tray If any parts to be cleaned are still attached to the Build Table ensure that the hanging arm is secured to th...

Страница 16: ...cycle Before starting the cleaning cycle ensure the following conditions are met The outer container inner container inner container plug hanging arm Build Table and lid are all secured in their prope...

Страница 17: ...ints may require more than one wash Do not continuously wash a part for more than 10 minutes as continuous and prolonged submersion in isopropyl can damage the print Recommended Clean Times minutes Re...

Страница 18: ...he 3D printed parts loaded in the B9Clean do not soak in isopropyl alcohol longer than the specified cleaning cycle duration the unit will automatically drain the inner container of isopropyl alcohol...

Страница 19: ...esin used and part geometry It is solely at the discretion of the operator to determine when the isopropyl alcohol is to be changed 9 2 2 Disposal Isopropyl alcohol is classified as a hazardous waste...

Страница 20: ...s plugs included AU EU UK US Required table area 300 mm x 300 mm Package size W x H x D 300 x 365 x 570 mm 11 8 x 14 4 x 22 4 Package weight 5 kg 11 lbs Agitation method Magnetic stirrer Cleaning solv...

Страница 21: ...in the table Error description Possible Causes Remedy 1 The indicator lights on the control interface are not lit The power adapter is not plugged into the back of the B9Clean base Plug the power ada...

Страница 22: ...cleaning cycle the isopropyl alcohol in the inner container is not being agitated The magnetic stirring bar is not placed in the bottom of the inner container Press the Start Stop button to stop the...

Страница 23: ...eaning cycle duration or spin speed settings are too low Increase the duration and spin speed settings to find the optimal values for a particular 3D printed part The 3D printed parts being cleaned ha...

Страница 24: ...n and spin speed of a cleaning cycle can be adjusted while the cleaning cycle is in Part height less than 70 millimeters mm Outer container filled to 1 3 liters Inner container with drain plug in top...

Страница 25: ...propyl alcohol should be replaced with a fresh batch if the current batch appears dirty or is not cleaning parts effectively 15 Document History Edition Date Comments released 031319 Updates by EH A 0...