3 General Description
©
b2 electronic
GmbH
DHV0078 b2 EN
13
Characteristic
1
HVA30-7
HVA40-5
HVA54-3
HVA68-2
Article number
SH5005
SH5009
SH5012
SH5016
Input supply voltage
190-240 V, 50/60 Hz
Input supply power
3 kVA
max. output
voltage
VLF sine wave
24 kV
rms
34 kV
peak
32 kV
rms
45 kV
peak
38 kV
rms
54 kV
peak
48 kV
rms
68 kV
peak
VLF square wave
34 kV
45 kV
54 kV
60 kV
DC [+/-]
34 kV
45 kV
54 kV
60 kV
resolution: 0.1 kV, accuracy: ± 1%
max. output current
120 mA
rms
52 mA
rms
resolution: 1 µA, accuracy: ± 1%
Resistance range
0.1 MΩ….5 GΩ
Output frequency
0.01 Hz-0.1 Hz in steps of 0.01 Hz (default 0.1 Hz) –
auto frequency selection
Sheath test
max. test voltage: 10 kV, duration: 1 min-15 min
trip current: 0.1 mA-5.0 mA
Sheath fault location mode
2
max. test voltage: 10 kV, duration: 1 min-60 min
pulse/period: 1:3 / 4 s, 1:5 / 4 s, 1:5 / 6 s, 1:9 / 6 s
Frequency optimization
Yes
Output Load
Capacity
at 0.1 Hz
7 µF
5 µF
3 µF
2 µF
at reduced
frequency/voltage
15 µF
15 µF
10 µF
10 µF
Metering
voltage and current (true rms and/or peak),
capacitance, resistance, time, flashover voltage
Output duty
Continuous! No thermal limitation for operating time.
Test modes
manual & automatic
Output modes
AC (VLF) symmetrical and load independent across full range,
DC (plus or negative polarity), burn/fault condition or
fault trip mode, jacket/sheath testing
Safety
50 Hz-12 kV feedback protection/dual discharge device (internal)
Computer interface
RS232 & USB flash memory module
Record storage
Built-in memory: up to 50 reports, 40 test sequences
USB memory flash drive: unlimited
PC software [included]
b2 ControlCenter for Windows
Weight
57 kg
Dimensions
4
(mm) L x W x H
450 x 340 x 520
Environment
Storage temp.
-25°C to 70°C
Operation temp.
-10°C to 50°C
Humidity
5-85% non-condensing
Содержание HVA34
Страница 2: ...1 Introduction 2 DHV0078 b2 EN b2 electronic GmbH...
Страница 11: ...2 Safety b2 electronic GmbH DHV0078 b2 EN 11...
Страница 22: ...4 Design and Construction 22 DHV0078 b2 EN b2 electronic GmbH Switches and controls Front panel HVA34...
Страница 33: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 33 Connection diagram vacuum bottle test...
Страница 47: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 47 5 5 Discharge Status...
Страница 62: ...11 Declaration of Conformity 62 DHV0078 b2 EN b2 electronic GmbH 11 Declaration of Conformity HVA34...
Страница 63: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 63 HVA30 5...
Страница 64: ...11 Declaration of Conformity 64 DHV0078 b2 EN b2 electronic GmbH HVA30 7 HVA40 5 HVA54 3...
Страница 65: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 65 HVA54 5...
Страница 66: ...11 Declaration of Conformity 66 DHV0078 b2 EN b2 electronic GmbH HVA60...
Страница 67: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 67 HVA68 2...
Страница 68: ...11 Declaration of Conformity 68 DHV0078 b2 EN b2 electronic GmbH HVA90 HVA94...
Страница 69: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 69 HVA120...