ULT FREEZER – USER MANUEL
41
Updated 19. April 2021
5. Power Alarm
Setting for Power Alarm.
1.
You can enable or disable the Power failure alarm. Push ON or OFF.
2.
To save the changed data click on the OK button.
In case of Power cut, the Power failure alarm will be activated.
In case of Power Failure, the display will turn black to save Battery power, and the
controller will give a beep every 1 minute.
To see the actual temperature, press your finger somewhere in the display shortly, and the
display will light up and show the Power failure alarm, and the actual temperature in the
chamber. After few seconds, the display will turn black again, just for saving battery power.