2.1.2  Cautions 

• Ensure that there is sufficient room around the unit for air circulation. A good air circulation especially

around the compressor is essential for trouble-free operation. Keep the ventilation openings of the
compressor cover always free from any obstruction. Make sure that a minimum distance is provided
between the unit and any wall or other device located next to it.

• Do not locate the unit below a ceiling fan or right next to air-conditioning equipment.
• The sound intensity level produced by the appliance is lower than 70 dBA  (measured at a distance of

1 meter).

• Check whether the appliance has been delivered undamaged. If you find that damage has occurred in

transit, immediately contact the delivery service or relevant sales outlet, submitting the delivery note or
proof of purchase. 

• Do not operate a unit that has been damaged in transit! If you are unsure, contact your sales outlet and

ask them.

• The device must be transported in an upright position only (maximum inclination 45°).
• All connected devices/signalling equipment must provide reinforced or double insulation for protection

against electric shock.

• Upon the occurrence of an alarm situation, attempts must be made to discover the reason for the alarm

and resolve it as quickly as possible. If that is not successful the necessary measures must immediately
be taken so that the stored goods are not damaged.

• Optimal function of the appliance is given at an ambient temperature indicated in the technical data (see

chapter 10) with a maximum relative humidity of 95%.

• The device must be set up in a dry well-ventilated place. Avoid direct sunlight or locating it close to a

heat source. The location must be protected against rain and dust.

• This device cannot be used/is not for use in facilities at altitudes of 2000m or higher.
• The use of hydrogen peroxide H2O2 (VHP) for the disinfection (sterilization) of the appliances is not

suitable. This procedure may be applied exclusively with dedicated appliances.

• Other than the cleaing procedure described in this manual, follow the internal policy and procedures.
• Before using the refrigerator it should be cleaned inside and outside.
• For cleaning, use only gentle cleaning agents. Never use aggressive or caustic cleaning agents, scouring

powder, steel wool, abrasive sponges or chemical solvents. When cleaning, make sure that no fluids of
any kind run into the ventilation housing.

• For disinfecting, we recommend all the surface disinfecting agents commonly used by the customer,

provided they are recommended by the national organizations. For disinfecting small areas, we
recommend using a concentrated alcoholic agent.

• Before being loaded, the unit must have reached the set point temperature.
• The load should only be stored in the provided baskets and should not exceed the storage areas. The

total amount of stored blood bags should be evenly distributed on the 5 baskets (max. 18 blood bags
per basket) (image 6). Do not overload the device.

• Different blood or plasma types should be sorted and stored in a manner for the User to identify these

with easy. 

• The loading must never obstruct the air-inlet-openings.
• Do not store warm items in the devices. They are not designed for fast cooling of warm items. 
• Avoid long and / or frequent lid openings to prevent that the inner temperature rises to high.
• Make sure that the lid is correctly closed by using the supplied latches and that the lid gasket seals

correctly against the cabinet. 

•  Do not bring more than 18 new blood bags per day into the MRB3000SDD.

Содержание MRB 3000 SDD

Страница 1: ...820 9506 46 ed0821 MRB 3000 SDD SOLAR DIRECT DRIVE ICE LINED BLOOD STORAGE REFRIGERATOR Instructions for Use...

Страница 2: ......

Страница 3: ...certified according to the safety standard IEC 60335 1 This equipment is designed to be safe at least under the following conditions based on the IEC 60335 1 Indoor use Altitude up to 2000 m Temperat...

Страница 4: ...________________15 4 2 Opening and closing___________________________________15 4 3 Important Advice for Storage ___________________________16 4 4 Operating controls __________________________________...

Страница 5: ...5 1 Model Overview Group Model name Ice lined Blood Storage Refrigerators MRB3000SDD...

Страница 6: ...refrigerator for Solar Direct Drive without battery storage t has a bank of icelining elements installed inside the cabinet During operation these elements freeze and act as cold storage for protectin...

Страница 7: ...n that on the nameplate may cause fire or electric shock Do not store flammable or explosive substances inside these appliances e g aerosols with flammable propellant These may cause explosion or fire...

Страница 8: ...source The location must be protected against rain and dust This device cannot be used is not for use in facilities at altitudes of 2000m or higher The use of hydrogen peroxide H2O2 VHP for the disinf...

Страница 9: ...bmitting the delivery note or proof of purchase Do not operate a unit that has been damaged in transit If you are unsure contact your sales outlet and ask them The unit must be transported in an uprig...

Страница 10: ...the ice lined blood storage refrigerators Patients do not come into direct contact with the refrigerators 2 4 3 Intended use environment The Ice lined blood storage refrigerators are stationary type...

Страница 11: ...between 2 C and 6 C for no longer than five days prior to use After concentration and irradiation the storage time must be no longer than 24 hours 2 4 5 Contraindications The devices are not intended...

Страница 12: ...ctric shock may result due to corrosion The wall socket must be easily accessible When the unit is in its final position it is important to immobilise the unit using the feet CAUTION Optimal function...

Страница 13: ...e described in this manual follow the internal policy and procedures Before using the refrigerator it should be cleaned inside and outside For cleaning use only gentle cleaning agents Never use aggres...

Страница 14: ...ms with other electrical equipment causing malfunction the unit should be connected to a separate circuit Never connect it to a single socket with other electrical equipment by means of a multi socket...

Страница 15: ...ights up when there is a minimum of solar energy available Supposed that there is enough solar energy available after switching on the device starts working after a delay of 5 minutes This also applie...

Страница 16: ...nner temperature rises to high Make sure that the lid is correctly closed by using the supplied latches and that the lid gasket seals correctly against the cabinet Do not bring more than 18 new blood...

Страница 17: ...re and the displayed max and min values 4 4 3 Display In normal use the inside temperature of the appliance is displayed in the main area of the diplay C with one decimal max min main area min area ma...

Страница 18: ...error E1 4 Warning Display sensor below reading range E1 5 Warning Display sensor above reading range E2 1 Alarm Heating sensor not connected E2 2 Alarm Heating sensor short circuit E2 3 Alarm Heating...

Страница 19: ...cessary for the correct daily use of the SDD unit its presence is mandatory When it is not connected an alarm is created When it is damaged detected by a low voltage on the poles another alarm is crea...

Страница 20: ...ositioning of solar panels etc Check the actual situation of the panels and if the problem persists after corrective actions notify Refrigerator Technician Thermal cut out of electronic unit E5 5 The...

Страница 21: ...the capacity of the battery is no longer sufficient to take over the control function the alarm message E4 2 appears on the display If no battery is connected the system will generate an E4 1 alarm me...

Страница 22: ...ly gentle cleaning agents Never use aggressive or caustic cleaning agents scouring powder steel wool abrasive sponges or chemical solvents When cleaning make sure that no fluids of any kind run into t...

Страница 23: ...side air to go in and form condensation In such cases the hinges should be adjusted or the gaskets should be changed Restrict and reduce frequency of opening the MRB3000SDD Open only when absolutely n...

Страница 24: ...values Check if the ambient temperature is too high outside of the opreating range of the device which can cause the interior temperature too rise Check if there is enough air circulation arround the...

Страница 25: ...perational C Max Humididty range Altitude Refrigerant Type Refrigerant quantity g Hold over time at 32 c Hold over time at 43 c Cool down Energy consumption stable running Energy consumption cool down...

Страница 26: ...26...

Страница 27: ...27...

Страница 28: ...B Medical Systems S r l op der Hei 17 L 9809 Hosingen Luxembourg...

Отзывы: