During a maximum transit time of 24 hours, the temperature of whole blood or other blood components
must at no time 10°C. Temperature excursions above +10°C increases the risk of sepsis due
to bacterial contamination.
Blood components intended for intrauterine and exchanged transfusion (e.g. IUT, ET, PR/ET) must be
stored between 2°C and 6°C for no longer than five days prior to use. After concentration and irradiation,
the storage time must be no longer than 24 hours.
2.4.5. Contraindications
The devices are not intended to be used for quick cooling of blood or blood components.
The devices are not intended to be used for the cooling of foodstuffs or any other products not covered
in the intended purpose.
The devices are not intended to be operated outside the environmental and installation conditions
documented in the operating instructions.