the upper alarm limit is reached. Triggering of the
temperature alarm.
2. Phased reduction of the internal temperature until
the lower alarm limit is reached. Triggering of the
temperature alarm.
q
REMOTE ACTIVATION
8
w
REMOTE OFF
pq
ON
8
The remote alarm function can be switched on and
off for an alarm test.
q
COMM. ADDRESS
8
w
COMM. ADDR: 0
pq
255
8
With the aid of the built-in RS485 interface several
appliances can be networked with each other. If this
is the case every individual appliance must receive a
separate communication address.
(setting of 0 – 255)
q
SERVICE
8
w
RESET SERVICE TIME
8
w
RESET COUNTER
8
w
YES
pq
NO
8
After the expiry of a given operating time a “SER-
VICE” message appears on the display to call atten-
tion to the fact that a regular maintenance is due.
The counter for this service time can be put back
with this function.
q
HIDE SERVICE WARNING
8
This function is for cancelling the “service”
message.
q
USER PASSWORD
8
Changing the password to access the “SETTINGS”
MENU.
w
CHANGE USER PASSWORD:
0XXX
XXXX
8
Enter the new password twice and confirm.
q
CALIBRATION
8
This function is for calibrating the sensors. As each
setting in the calibration compromises the function of
the appliance, this function is protected by a pass-
word in the factory and must only be used by trained
service staff!
w
CALIBRATION PASSWORD:
0XXX
8
Enter the password to the calibration menu
5.4 The “EXTRA” - Menu
• The “EXTRA” menu only contains parameters without influence on the function of the appliance and are
therefore accessible for any use without a password.
3
press the MENÜ-key
q
select
“EXTRA”
8
w
HISTORY
8
(see chapter “Alarm history” for further details)
w
SHOW HISTORY
8
Display for alarm situations. With the aid of the UP
and DOWN buttons you can scroll down the list
q
DELETE HISTORY
8
w
DELETE HISTORY:
NO
pq
YES
8
YES : deletes the entire history data.
q
BUZZER
8
w
VOLUME
8
21
Содержание MB 3000 G
Страница 1: ...820 9505 61 ed0821 Instructions for Use MB 3000 G ICE LINED BLOOD STORAGE REFRIGERATOR...
Страница 2: ...2...
Страница 3: ...3...
Страница 6: ...6 1 Model Overview Group Model name Ice lined Blood storage Refrigerators MB3000G...
Страница 31: ...31...
Страница 32: ...B Medical Systemes S r l op der Hei 17 L 9809 Hosingen Luxembourg...