- 13 -
16. Do not connect
Stimuplex
®
HNS 12 to other devices (except for instruments and acces-
sories, see section 5.10). I
f Stimuplex
®
HNS 12 is operated near another device, the user
must monitor the equipment or system and check that the configuration used in this way
is operated properly and as intended.
17. Other devices can interfere with Stimuplex
®
HNS 12 or the system, even if they show
compliance with the applicable CISPR emission requirements. Interference impulses can
be picked up by the stimulation current detector and thereby trigger an error display and,
possibly, a safety shutoff.
18. Operation of other devices or systems with Stimuplex
®
HNS 12 accessories can increase
emissions or reduce immunity to interference on devices or systems. Observe the sup-
plied EMC instructions concerning installation, initial setup and operation of the device
or system (see Section 9. Electromagnetic compatibility (EMC)).
19. Do not use the device after ingress of liquids. Ingress of liquids or humidity can cause an
electronic failure.
20. To avoid damage to the connecting cable and the device, do not hold or carry the device
by its connecting cables and/or its accessories. Do not wrap the cable around the device
or around other equipment.
21. Wrapping the connecting cable during normal operation of the stimulator generates
inductive components and, at very short stimulation pulses, can lead to a reduction in
stimulation efficacy and/or false measurements of the actual stimulation current. False
interpretations of the indicated values can be the result.
22. For safety reasons, never operate Stimuplex
®
HNS 12 if the battery is leaking, but return
it to the manufacturer for proper cleaning. The same applies if any liquid soaks through!
23. Do not continue to operate the device if repeated error messages appear. The device must
be returned to the manufacturer for proper repair.
24. Known undesirable side effects: Hematoma at puncture site
Содержание Stimuplex HNS 12
Страница 1: ...Stimuplex HNS 12 Nerve Stimulator for Peripheral Regional Anesthesia User Manual GB ...
Страница 19: ... 19 Overview of the Stimuplex HNS 12 Nerve Stimulator ...
Страница 65: ... 65 Target stimulation current as a function of adjustment control setting relative to 0 Setting 2 turns ...
Страница 66: ... 66 Appendix C according to IEC 60601 2 10 ...
Страница 77: ......