- 45 -
4.2.7.3. Current adjustment
This option allows automatic adjustment of the set stimulation current to the new settings
when the stimulus duration is increased
during an application
. This prevents the set stimu-
lation current from suddenly being given off for too long when the stimulus duration is
increased without the stimulation current having been turned back to 0.00 mA.
When in SENSe mode, there is no automatic current adjustment, since the stimulus duration
is fixed and can not be changed via the quick access button for impulse duration „ms“.
When the stimulus duration is lowered with the ms key and then increased
again without the dial being operated in the interim, the target stimulation
current is turned up to the old value again.
In the ”Setup“ menu, open ”Options”, highlight ”Auto adjust current” and activate (
) or
deactivate with the right arrow key.
4.2.7.4. Additional stimulus duration
With this option, you can add the stimulus durations 0.05 ms and 0.50 ms to the selection
series (see Section 3.4). In ”Setup”, open ”Options“, highlight ”Add. stimulus duration” and
activate (
) or deactivate with the right arrow key.
4.2.7.5. SENSe option
With this option enabled, you can add the SENSe stimulation mode (see Section 2.2.) to the
2 Hz/1 Hz series. In „Setup“, open „Options“, highlight „SENSe“ and activate (
) or deacti-
vate with the right arrow key.
Содержание Stimuplex HNS 12
Страница 1: ...Stimuplex HNS 12 Nerve Stimulator for Peripheral Regional Anesthesia User Manual GB ...
Страница 19: ... 19 Overview of the Stimuplex HNS 12 Nerve Stimulator ...
Страница 65: ... 65 Target stimulation current as a function of adjustment control setting relative to 0 Setting 2 turns ...
Страница 66: ... 66 Appendix C according to IEC 60601 2 10 ...
Страница 77: ......