44
Technical data
Parameter
Value
Interfaces
Q
Cold connector for mains voltage
Q
Accessory port for interface cable 12 V CP and staff call
Q
IrDA infrared for communication in the station and for
service
Operating conditions
Q
Temperature
Q
Relative air humidity
Q
Atmospheric pressure
Q
+10 °C … +40 °C / +50 °F ... +104 °F
Q
30% … 90% (without condensation)
Q
0.54 … 1.06 bar
Storage conditions
Q
Temperature
Q
Relative air humidity
Q
Atmospheric pressure
Q
–20 °C … +55 °C / -4 °F ... +131 °F
Q
20% … 90% (without condensation)
Q
0.5 … 1.06 bar
Weight
Approx. 1.9 kg
Dimensions in mm
(W x H x D)
Approx. 229 mm x 98 mm x 220 mm
(including compact
plus
)
Safety check
Every 2 years
Volume preselection
0.1 ml - 9,999 ml in increments of 0.01 ml
Time preselection
00:01 h - 99:59 h
Delivery accuracy
±5% according to IEC/EN 60601-2-24
Note:
valid for 50 cm water column
Occlusion alarm pressure 9 levels up to 1 bar
Alarm in the case of
incorrect dose
In the event of an incorrect dose of 1 ml due to pump
malfunction (electronics, software), the pump will automati-
cally switch off.
Delivery rate increments
0.1 ml/h ... 1200 ml/h in increments of 0.01 ml/h
Delivery accuracy for
bolus administration
typ. ± 5% for bolus volumes > 1 ml
Max. bolus volume after
bolus reduction
≤
0.2 ml