49
Electromagnetic compatibility
Immunity tests
Test level
EN 60601-1-2
EN 60601-2-24
Compliance
level
Electromagnetic
environment guidelines
Voltage dips,
brief supply volt-
age interruptions
and fluctuations
according to
IEC 61000-4-11
<5% UT ¹
for ½ periods
(>95% dip)
Complies
through
the use of
an internal
energy source
The supply voltage quality
should be the same as that
of a typical commercial or
hospital environment.
40% UT ¹
for 5 periods
(60% decline)
70% UT ¹
for 25 periods
(30% decline)
<5% UT ¹
for 5 s
(>95% dip)
Magnetic field at
supply frequency
(50/60 Hz)
according to
IEC 61000-4-8
3 A/m
400 A/m
Magnetic fields at the
supply frequency should
correspond to those typi-
cally found in commercial
and hospital environments.
Conducted HF
interference
according to
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
Outside ISM bands
10 V
rms
In all bands
Do not use portable and
mobile radio communica-
tions equipment closer to
the Infusomat compact
plus
(including connection
cables) than the recom-
mended safe distance calcu-
lated using the appropriate
equation for that frequency.
Recommended safe distance:
d = 1,2 √P³
10 V
rms
Within ISM bands