Revision Service-Documentation

Version 2.1

This Service Manual was approved by B. Braun on 16.03.2006. 

This manual has been completely revised. The most important 

changes are listed below:

-

Changed manual structure

-

New software

-

New spare parts

-

Total list of spare parts

-

Checking the pressure cut-off during Bolus after repair

Current Information

If you hear a scraping noise when the drive arm is pulled out, the 

straight pin (under the spindle) may have come loose. In this case, 

an additional straight pin lock (Ord. No. 3450 9100) can be insert-

ed in units up to serial No. 66725.    From serial No. 66725 on this 

straight pin lock is already fitted. Observe the instructions at-

tached.

A - 2

Содержание 0871 4827

Страница 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10 ...

Страница 2: ...h 8713 9112 Languages of this Manual The Service Manual for this unit can be supplied in the following languages Designation Part No Service Manual Perfusor compact german 8713 9111 The complete Service Manual contains the following pages Page 0 1 to page 0 10 Page 1 1 to page 1 4 Page 2 1 to page 2 6 Page 3 1 to page 3 16 Page 4 1 to page 4 18 Page 5 1 to page 5 6 Page 6 1 to page 6 2 Page 7 1 to...

Страница 3: ...ge 1 1 Function Page 1 2 Accessories Page 1 3 Software Approved Software Versions Page 2 1 Version Display during Switch On Test Page 2 1 Extended Version Display during Switch On Test Page 2 1 Error Messages and Alarms Page 2 2 Service Program Software Compatibility Page 3 1 Introduction Page 3 1 Working with the Service Program Page 3 3 What to Do if Trouble Shooting Page 3 5 Menu Description Pa...

Страница 4: ... Electrical Safety Page 5 5 Maintenance Page 6 1 Technical Safety Check TSC Page 7 1 Procedural Instructions on the TSC Visual Inspection Page 8 1 Functional Inspection Page 8 2 Pressure Cut Off Page 8 3 Syringes Page 8 4 Electrical Safety Page 8 4 Accessories Page 8 4 Optional Page 8 4 Test Equipment and Special Tools Page 9 1 Spare Parts List Page 10 1 Index Page 11 1 Appendix Revision Service D...

Страница 5: ... the pe riods stated 6 MP BetreibV B Braun also recommends training on the Technical Safety Checks or to perform at least the steps indicated in the current version of the manual as the TSC requires that the instructions in the manuals are ob served the manuals are a reference for measurements depending on the unit type the Service Program must be called which may lead to a dangerous unit conditio...

Страница 6: ...sion service the technician has participated in a technical training course for the specific B Braun unit Quality Management B Braun is certified in accordance with DIN EN ISO 9001 and ISO 13485 This certification also includes maintenance and serv ice The unit has the CE label The CE label confirms that the device corresponds to the Directive of the Council for Medical Products 93 42 EC of June 1...

Страница 7: ...trically conductive shoes to prevent electrostatic charge Spare Parts and Test Equipment Only use original spare parts from the manufacturer Do not tamper with assembly groups which can only be exchanged com pletely The spare parts required are listed in Section 9 Service personnel are responsible for the calibration of their test equipment Original test equipment can be calibrated at the works of...

Страница 8: ... are displayed green Click with the mouse button on a reference to jump to the corresponding source Menu commands are described as Menu File List of Abbreviations Abbreviations which are not generally known but are used in this manual are listed below A Module Analog Module E Module Electronic Module ESD Electrostatic Discharge IfU Instructions for Use LCD Liquid Crystal Display MFC Multi Function...

Страница 9: ...ternational Technicians Intercompany Nadja Machal Fax 49 5661 75 47 89 e mail nadja machal bbraun com Service Hotline Karl Tippel Tanja Kördel Phone 49 5661 71 35 25 Fax 49 5661 71 35 26 e mail karl tippel bbraun com e mail tanja koerdel bbraun com Return of Spare Parts and Test Equipment B Braun Melsungen AG Schwarzenberger Weg 73 79 Wareneingang Werk C 34 212 Melsungen Germany Safety Officer 30 ...

Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes ...

Страница 11: ... the membrane keyboard The syringes are changed semi automatically the function process and moni toring is microprocessor controlled The Perfusor compact has a long service life and is easy to service due to its modular design Individual modules can be replaced easily and quickly and the Service Program runs on a PC Fig 1 1 Carrying handle Membrane keyboard MFC socket LC display Syringe holder Axi...

Страница 12: ...nterface 2 E Module as operating and control unit 3 Drive unit consisting of drive board with the complete sensor technology light barriers for syringe pre and end alarm syringe size rec ognition and motor operation control push button sensor board for the inserted syringe positive locking sensor board for the frictional connec tion between nut and spindle of the drive Fig 1 2 Block diagram ...

Страница 13: ...ew 1 Perfusor compact 2 1 gb 1 3 Accessories Designation Ord No Unit connecting lead 200 240 V 3450 2718 Unit connecting lead 100 120 V 3450 5423 Pole clamp universal clamp rotating 3450 9054 Battery pack 3450 1690 ...

Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes ...

Страница 15: ...he other on screen 888 8 111 1 222 2 555 5 AA Reference to the instructions for use hard and soft ware group 3 The Perfusor compact switches over to normal operation Extended Version Display during Switch On Test 1 Switch on unit 2 Press the F button and keep the button pressed during normal switch on test The following information examples appears on screen after the information displayed during ...

Страница 16: ...nd CMP condition 3 Rate of FP and CMP different 4 Different function mode 5 Different rate of delivery 6 Different target volume 7 Different step volume low 8 Different motor steps 19 State motor state 20 Invalid normal state 21 return from PlcMain 22 Unexpected reset 28 No sync at Plc_Down 29 No sync at Plc_On 30 Different CMP FP mode ports 31 Invalid mode ports 32 Invalid variable values 33 Erro...

Страница 17: ...ce between NEC H8 85 Bw difference between NEC H8 86 Md difference between NEC H8 90 Syringe state in Oper_Syr 91 Set syringe type 92 Consistency error 93 Difference between setting and display 94 Timer synchronization 100 Division by zero 101 Illegal zero pointer 102 Illegal switch to default 105 No contact to NEC in OFF 110 Alarm on CMP side 111 119 Motor test 1 9 120 Motor current flow in OFF 1...

Страница 18: ...16 150 Different software versions 151 Double CRC error 153 Different states 154 Different rates 155 Different F mode 156 Different mode values 157 Different alarm recognition 158 Different alarm clearance 159 Err current volume 160 Err preselected volume 161 Err volume per step 170 Sensor sync failed 171 174 Sensor dark test error 180 ROM test error 181 RAM test error 182 Keyboard test error colu...

Страница 19: ... at Plc_Down 208 No sync at Plc_On 209 CMP FP timer end sync error 220 Different phases busy 221 Different phases idle 222 Motor on at reverse steps 223 Too many pending steps 224 Motor current error 225 Error of motor step number 226 Reverse polarity of motor 227 Motor steps overflow 230 Different syringe recognition 231 CMP FP syringe state 232 CMP FP syringe type set 241 249 Motor test 1 9 erro...

Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes ...

Страница 21: ...cription refers to the Service Program version 6 0001 Introduction The Service Program runs on a PC All functions are easy to oper ate in the pulldown menus as in Windows The special keys on the keyboard have the following functions ESC Escape F1 Start F2 Default EEPROM F3 Read out EEPROM F4 Serial number F5 Length calibration F6 Pressure calibration F7 Write EEPROM F8 Save Alt A Alternatively Alt...

Страница 22: ...wing functions can be executed Drive calibration Reading loading pump data Displaying operation values Displaying and changing parameters Documentation of pump hardware modifications Saving all data to a diskette hard disk or similar Installation 1 Insert diskette 2 Start the File Manager or Windows Explorer 3 Start Setup exe with a double click The directory C PLC_SERV is created automatically by...

Страница 23: ...e 4 Call menu Modes Modification and menu Syringes Sy ringe selection or Syringe types Note down parameters pri or to any modification e g new E Module Adjust Unit Settings 1 Menu File Connect 2 Menu EEPROM Read 3 Desired modifications display please see Operation Operation values Modes Modification data Calibration Pressure calibration required in case of bolus rate change Syringes Syringe select...

Страница 24: ... changed Defaut Data The Service Program contains the Default dat file with the factory settings of the unit These values can be adjusted via the Syringe or Modes menu if required Max delivery rate basal rate 99 9 ml h Bolus rate 800 ml h Staff call dynamic with Off alarm Alarm tone 0 3 Hz interval tone Pressure stage 3 Dianet address 1 Syringe selection Table Service interval 20440 hrs ...

Страница 25: ...t Termination Switch off pump Repeat communication start Switch pump on again the communication cannot be started Was the setting in the File Configuration file COM 1 oder 2 menu selected correctly Is the service cable connection okay Is the MFC correctly connected the communication starts and is then interrupted Then Press the ON key on the Perfusor compact until the sym bols and disappear the un...

Страница 26: ...ion file 4 Exit Exits the Service Program EEPROM Menu 1 Read F3 The data of the Perfusor compact can be checked and mod ified in the Service Program after they were read out Note The hyphens after the menu items SNr DAbg and LAbg in the footer are deleted and replaced by the figure 1 This means that calibration data is now transmitted to the program 2 Default F2 Loads data of the Default file in t...

Страница 27: ...rriers and the drive end is determined by length calibration The motor steps deter mined are displayed after calibration is terminated 3 Pressure Calibration F6 The motor parameters for setting the 3 pressure stages and the correct switch off in bolus mode is determined by pres sure calibration 4 Overload Check see Overload Check p 3 15 The dynamic pressure test is used to determine whether the un...

Страница 28: ...usor compact Table Allows to select all syringe types saved in the Perfusor compact Free type Only the free syringe types can be used can be loaded Table and free type Allows to select all syringe types which are saved and can be loaded in the Perfusor compact OPS Defines that only OPS 50 ml and OPS 20 ml syringes can be used Fig 3 8 Fig 3 9 ...

Страница 29: ... file with zero value syringes Now the corresponding free syringe type is de leted If necessary adapt the syringe selection e g when the selection was set to Free type but all free types were deleted Note Adapt the syringe table after all modifications were carried out see Syringe Table and and Quick Reference Guide p 4 5 3 ROM 20 ml ROM 50 ml table Display of the syringe types saved in the Perfus...

Страница 30: ...bolus rate and the syringe selection were changed and the Service Program is quit CAUTION A pressure calibration must be carried out if the bolus rate was changed Constants Menu 1 Service interval Reads and resets the service interval timer When the time set has elapsed a service interval alarm is triggered when the unit is switched on The timer can be set to 20440 hours corresponds to an aver age...

Страница 31: ... Acknowledge by starting again Now the unit delivers the maximum basal rate Bolus Rate if not blocked 1 Press the F1 key bolus rate The value displayed must corre spond to the bolus rate set 2 Start bolus press F and 1 simultaneously Pump must deliver in bolus mode and the volume infused in bolus mode is dis played Staff Call 1 Plug MFC service connector on the MFC connector of the unit 2 Open syr...

Страница 32: ...its with unit software up to PLAA00070 0 Service Program version 6 001 Continuous tone with 3 Hz intermittent Continuous tone static Syringe Syringe Selection Note Note down the current code for 20 ml and 50 ml syringes before starting the test Test for table or internal setting 1 Open holder then press keys 7 C X X F and 7 C Y Y F The syringe is accepted if an alarm is triggered then the sy ringe...

Страница 33: ...etting is not correct 5 Open holder press keys 7 C number of the 20 ml free type F and keys 7 C number of the 50 ml free type F syringe is accepted if an alarm is triggered then the syringe selection setting is not correct Test for OPS setting 6 Open holder press keys 7 C 5 0 F and keys 7 C 2 0 F syringe is accepted if an alarm is triggered then the syringe selec tion setting is not correct 7 Pres...

Страница 34: ...enu File Connect F1 2 After activities on the drive Transfer data via menu EEPROM Read F3 from the unit to PC 3 After replacement of the E Module Call in default data from the PC via menu EEPROM De fault F2 The existing values are deleted 4 Check or input the serial number in menu Calibration Se rial Number Default data cannot be transferred to the device if the serial number was not input 5 Calib...

Страница 35: ...ural instructions for in spection see Procedural Instructions for Inspection after Modifications via the Service Program p 3 11 Overload Check 1 Connect 50 ml OPS syringe filled with water drawn up to 25 to 30 ml via infusion line to vented pressure gauge 2 Insert syringe and start overload check via menu Calibration Overload Test Overload check is started with a force of 50 and can be modified in...

Страница 36: ... shipped Test ok Calibration Serial number Modes max basal rate Modes Bolus rate Modes Staff call dynamic static Off alarm dynamic static Off alarm Modes Alarm tone Syringes Syringe selection Table Free type Table free type OPS Table Free type Table free type OPS Syringes Free 20 ml type Free 50 ml type Syringe type number Syringes Syringe selection as de livered 20 ml 50 ml ...

Страница 37: ...played upon startup Press the ON OFF button to delete the alarm 022 until the alarm symbol is no longer displayed If the alarm 105 is triggered afterwards switch the unit off Before startup 4 If batteries are used switch the device first on without mains connection If the battery pack is used then the device is to be switched on with mains connection Note Defective batteries must be disposed of ac...

Страница 38: ...e guide 5 screw split rivet for battery compartment cover 5 blind plug for syringe holder 5 countersunk screw M 3x10 5 flat head screw M 3x5 10 flat head screw M 3x6 5 board holder 5 flat head screw M 3 x 14 10 countersunk screw M 4x12 25 Ejot KM 22x8 15 tamper proof cap Unit connecting lead hospital grade 3450 5458 Unit connecting lead 220 240 V 3450 2718 ...

Страница 39: ...n the E Module when disconnecting 4 Dismount both housing halves Always check A Module before replacing the board Other modules can only be exchanged without danger of conse quential damage if there is no overvoltage Connect mains cable when the housing is open Measure voltage parallel with capacitor C3 The set value is 6 2 to 6 8 volt Fig 4 1 Ribbon cable to drive board Motor cable Fig 4 2 A Modu...

Страница 40: ...Checks after Repair Procedural instructions see Procedural Instructions for Inspec tion after Modifications via the Service Program p 3 11 A calibration in the Service Program is to be carried out if a new E Module is installed or the drive is replaced see Service Pro gram p 3 1 Fig 4 3 A Module board from serial number 38101 on ...

Страница 41: ...91 1329 Spanish 3891 1337 Dutch 3891 1345 Italian 3891 1353 Portuguese 3891 1361 Norwegian 3891 1370 Finnish 3891 1388 Swedish 3891 1396 Czech 3891 1400 Turkish 3891 1418 Polish 3891 1426 Danish 3891 1434 Exchange 1 Remove split rivet First pull up the head then remove rivet completely 2 Insert new syringe table and quick reference guide Fig 4 4 Syringe table Quick reference Split rivet guide ...

Страница 42: ...ge holder with pin punch 3 Remove screw 4 Pull off holder 5 Push spacer washer 0 8x4x3 2 on the shaft if the syringe is not recognized 6 Insert new syringe holder 7 Fit new screw not the old one and safety lock with Loctite 274 8 Replace new cap 4 4 Unit Feet Designation Ord No Unit feet 3450 6640 Note The feet can be turned and used once again Pull feet out and turn around or exchange Fig 4 5 ...

Страница 43: ...at the battery compartment cover does not get jammed Check for tight fit The battery compartment cover is also the holder plate for the pole fixation 4 6 Snap in Clip Designation Ord No Snap in clip and snap in lever 3450 6616 Exchange 1 Loosen 5 screws from the bottom and carefully open housing pay attention to the cable lengths 2 Exchange snap in clip and snap in lever 3 Close the unit Note Do n...

Страница 44: ...outside and press MFC socket inwards 3 Pull off the N Module connector slightly pull out the A Module 4 Pull off connector on the E Module 5 Replace A Module and check snap in hook on the board 6 Assembly is done in reverse order Connect mains connector correctly to the A Module Do not squeeze the cable see Close Unit p 4 4 Note The connector on the E Module can be easily connected when the E Modu...

Страница 45: ...1 Open battery compartment see Battery Compartment Co ver p 4 7 2 Remove batteries 3 Connect unit to mains and switch unit on 4 Disconnect from mains for a short moment when the switch on test is finished Pull mains connector and plug in again to trigger a device alarm code 22 continuous tone 5 Put a small flat blade screw driver carefully through the battery compartment opening and set the volume...

Страница 46: ...lay unlock zero force connector and lay ribbon cable 6 Connect connection cable 7 Insert new E Module at the side into the guide and position behind the holder Caution Do not damage the components 8 Push board in the guide to the right and insert a new board holder must engage in hole 9 Push ribbon cable in zero force connector until stop and lock on both sides can get jammed lock both sides 10 Cl...

Страница 47: ...er Note Lay two wire cable with mains connector behind bearing Con nect mains connector correctly to the A Module please see fig ure Do not squeeze the cable see Close Unit p 4 4 Note The connector on the E Module can be easily connected when the E Module is swivelled out see E Module p 4 10 4 11 Housing Upper Part Complete Designation Ord No Housing upper part incl membrane keyboard 3450 6586 car...

Страница 48: ...450 6624 Straight pin lock 3450 9100 WARNING THE DRIVE CONSISTS OF SAFETY RELEVANT PARTS OPERATIONAL RELIABILITY CAN ONLY BE GUARANTEED WHEN THE DRIVE IS EX CHANGED COMPLETELY Exchange 1 Open unit see Open unit p 4 3 2 Move drive arm to middle position and lock 3 Loosen screw on syringe size board spread snap in hook and remove board 4 Loosen both screws on drive and remove drive 5 Insert syringe ...

Страница 49: ...change 1 Open unit see Open unit p 4 3 2 Move drive arm to middle position and lock 3 Loosen both screws on drive 4 Lift drive until the axial positioner is free 5 Remove axial positioner by forcing apart Replace new axial positioner and make sure that the scraper ring is correctly fit ted 6 Assembly is done in reverse order Note Do not squeeze the cable see Close Unit p 4 4 7 Calibration is requi...

Страница 50: ...ile and slide until stopper of the aluminium profile from the motor side CAUTION Cable layout according to figure 7 Press board against stopper when screwing down Tighten screws hand tight 8 Fix satellite board Cable layout please see fig pc_20 Lay motor cable under the direction of rotation board before fixing the board Make sure that the slotted disk can turn freely and smoothly 9 Insert ribbon ...

Страница 51: ...t of Drive p 7 1 4 16 Drive Head Designation Ord No Drive head complete 3450 1720 Toggle 3450 1711 Exchange 1 Open toggle pull out drive head and close toggle again 2 Pierce tamper proof caps in drive head with a pointed screw driver and remove caps Fig 4 18 Fig 4 19 Drive head with lock and push button sensor ...

Страница 52: ...plug con nector 8 Push drive head housing towards the unit 9 Pull jaws off the drive tube 10 Pull drive head housing together with clamp off the drive tube 11 Assembly is done in reverse order Note Pay attention to cable layout The cable must be laid between the pins and under the light barrier Fig 4 20 Drive head housing Cap for drive head housing Release shaft Toggle Pressure spring Lever Cam sw...

Страница 53: ...therwise the housing is damaged b Clean adhesive position on new housing and stick type plate New type plates can only be ordered as spare parts if the old type plates are returned to B Braun see Orde ring of Spare Parts and Test Equipment p 0 9 3 Remove drive from old housing see Drive p 4 12 4 Install drive in new housing see Drive p 4 12 5 Close housing Note Do not squeeze the cable see Close U...

Страница 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes ...

Страница 55: ...positioner drive head Syringe table quick reference guide Membrane keyboard Battery compartment cover battery compartment and contacts Unit feet MFC connector Holder for pole fixation side snap in mechanism Mains lead Mains voltage acc to TSC ______ V Protective conductor resistance acc to TSC ______ Ω Patient leakage current acc to TSC _____ µA Mechanical inspection Holder for pole fixation Stack...

Страница 56: ...Inspection Mechanical Inspection 1 Check function of the holder for pole fixation 2 Check stacking function of the unit with respect to other units 3 Check function of the syringe holder with syringe 4 Check function of the drive head lock Switch on unit 1 Switch on unit and keep ON button pressed for max 20 sec Check the screen display during this time A device alarm is triggered if the button is...

Страница 57: ...f 12 3 ml h key sequence 1 2 3 The pump delivers The delivery rate set must be displayed 5 Open lock Alarm by buzzer and positive locking sensor alarm Drive stops Lock must not engage in upper position 6 Plug in MFC service connector and actuate the START button Drive delivers at 12 3 ml h 7 Change delivery rate to 96 ml h key sequence C 9 6 F during infusion The pump delivers The delivery rate se...

Страница 58: ...pecified pressure is reached 6 Enter value in check list 7 Set pressure stage 3 key sequence F 3 C 3 F START 8 Pump continues to deliver at 96 0 ml h and switches off when the specified pressure is reached 9 Enter value in check list WARNING DANGER OF INJURY DO NOT OPEN TOGGLE AS CALIBRATION GAUGE CAN RELEASE SUDDENLY 10 Release calibration gauge key sequence F 3 0 with service connector Syringe R...

Страница 59: ... syringe 6 Confirm syringe type 20 with F button 7 Press START button Pump delivers and triggers pre alarm at a volume of 5 ml 8 Bolus press buttons F and 1 simultaneously and keep buttons pressed until end alarm pump must not trigger pressure alarm Electrical Safety 1 Measure mains voltage and note down 2 Measure protective conductor resistance and note down 3 Measure patient leakage current as d...

Страница 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes ...

Страница 61: ...1 gb 6 1 6 6 Maintenance The unit is maintenance free A Technical Safety Check TSC see Technical Safety Check TSC p 7 1 is to be carried out every 24 months to check the op erational capability of the Perfusor compact 6 2 ...

Страница 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes ...

Страница 63: ...de snap in mechanism 2 Functional Inspection Switch on unit Compare with instructions for use LCD self test and audible alarm Compare set delivery date and value displayed Check switching capability of staff call accessories Switch on unit in battery mode and check self test Note If Battery discharged is dis played Charge battery or replace bat teries and repeat test Compare status display 000 A o...

Страница 64: ...short circuited mains inlet and plus pole in battery compartment top right 6 Accessories Enter MFC battery etc 7 Optional Rate limitation ml h Bolus rate limitation ml h Infusion line used for Technical Safety Check Type ___________________Manufacturer ___________________ Inspection performed by Test result Defects found which could endanger patients users or third parties Yes No Measures to be ta...

Страница 65: ...stening Check function with OPS 50 ml syringe Syringe holder axial positioner drive head clamp and push but ton sensor Membrane Keyboard Check adhesion cleanliness and fit Battery Compartment and Contacts Check state of contacts tight fit not bent Unit Feet Check unit feet for completeness and proper fit Mains Lead and Connector Completeness damage MFC Lead and Connector Completeness damage Holder...

Страница 66: ...LC display if a syringe was not inserted 5 Insert current step gauge in the OPS 50 ml slot and close sy ringe holder The lock must engage automatically and the syringe symbol in the LC display must not flash 6 Confirm current step gauge as OPS with F button Program as OPS syringe type beforehand if necessary 7 Start pump with a delivery rate of 12 3 ml h key sequence 1 2 3 The pump delivers The de...

Страница 67: ... limit values of the Checklist for Checks after Repair do not correspond to the TSC values 2 Pump continues to deliver at 96 0 ml h and switches off when the specified pressure is reached 3 Enter value in check list 4 Set pressure stage 2 key sequence F 3 C 2 F START 5 Pump continues to deliver at 96 0 ml h and switches off when the specified pressure is reached 6 Enter value in check list 7 Set p...

Страница 68: ...move potentiometer calibration gauge c Insert max 50 ml potentiometer calibration gauges The 50 ml syringe symbol is displayed d Remove potentiometer calibration gauge Electrical Safety 1 Measure mains voltage and note down 2 Measure protective conductor resistance and note down 3 Measure patient leakage current as described hereafter and note down Remove battery pack or batteries Unit is at rest ...

Страница 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb ...

Страница 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes ...

Страница 71: ...50 1703 Calibration gauge 6 6 80 N 0770 1535 Current step gauge 10 130 N 0770 1616 for TSC as option Potentiometer calibration gauge 20 ml 0770 1543 Potentiometer calibration gauge max 50 ml 0770 1624 MFC service connector 3450 1215 Socket spanner for MFC connector 0770 1497 Interface cable 0871 1661 9 2 ...

Страница 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes ...

Страница 73: ...388 Swedish 3891 1396 Czech 3891 1400 Turkish 3891 1418 Polish 3891 1426 Danish 3891 1434 Syringe holder complete 3450 6608 with screw and cap Unit feet 3450 6640 Battery compartment cover 3450 6632 Snap in clip and snap in lever 3450 6616 A Module complete with board MFC and buzzer 3450 5288 replaces A Module up to serial number 38100 Buzzer 3450 8643 LS clip 3450 7710 E Module with DIANET 3450 6...

Страница 74: ...3450 6624 Straight pin lock 3450 9100 Axial positioner 3450 6659 Drive board 3450 6691 with main PCB and satellite boards for syringe size recognition and recognition of direction of rotation Drive head complete 3450 1720 Toggle 3450 1711 Housing bottom part complete 3450 6594 ...

Страница 75: ... hook 3450 8368 Plate 2 pcs 3450 2610 Connection cap D12 4 5 pcs 3477 4149 Bellows 5 pcs 3477 3274 Pressure spring 5 pcs 3477 4165 Universal Clamp Universal clamp complete not available any more Threaded rod 34 50 5903 Safety hook 34 50 5865 Turning handle 34 50 5890 Rubber cover 5 pcs 34 77 3290 Bellows 5 pcs 34 77 3274 Connection cap 5 pcs 34 77 3304 Pressure spring for pole fixation 5 pcs 34 77...

Страница 76: ...10 4 Perfusor compact 2 1 gb For your notes ...

Страница 77: ...s 2 2 Ersatzteilliste 10 1 F Function 1 2 Functional inspection 5 2 8 2 H Housing bottom part 4 17 Housing upper part 4 11 L List of abbreviations 0 8 LS clip 4 9 N N Module 4 11 O Operation 5 3 Overload test 3 15 P Physical construction 1 1 Pre alarm 5 5 Pressure cut off 5 4 8 3 Q Quick reference guide 4 5 R Repair information 4 1 Responsibilities 0 6 Revision service 0 5 S Service Program 3 1 In...

Страница 78: ...Index 11 11 2 Perfusor compact 2 1 gb Test equipment 9 1 Trouble shooting 3 5 TSC 0 5 7 1 U Unit feet 4 6 V Version display 2 1 Visual inspection 5 2 8 1 ...

Страница 79: ...software New spare parts Total list of spare parts Checking the pressure cut off during Bolus after repair Current Information If you hear a scraping noise when the drive arm is pulled out the straight pin under the spindle may have come loose In this case an additional straight pin lock Ord No 3450 9100 can be insert ed in units up to serial No 66725 From serial No 66725 on this straight pin lock...

Страница 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes ...

Страница 81: ...i cult This modification is limited to installation of a washer Modification 1 Pierce through the cap and remove 2 Fasten syringe holder with pin punch 3 Remove screw 4 Pull off holder 5 Push washer 0 8x4x3 2 on the shaft 6 Insert syringe holder 7 Fit new screw not the old one and safety lock with Loctite 274 8 Place on cap 9 Check syringe recognition for proper function see Functio nal Inspection...

Страница 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes ...

Отзывы: