Unit Elements
4
4 - 4
Perfusor® compact, 2.1 gb
Close Unit
1.
Close unit in reverse order of opening.
Note
Do not squeeze motor cable.
Checks after Repair
Procedural instructions
(see „Procedural Instructions for Inspec-
tion after Modifications via the Service Program“
A calibration in the Service Program is to be carried out if a new
E-Module is installed or the drive is replaced
Fig.: 4 - 3 A-Module board (from serial number 38101 on)