Unit Elements
4
Perfusor® compact, 2.1 gb
4 - 17
4.17 Housing Bottom Part, Complete
Designation
Ord. No.
Housing bottom part, complete . . . . . . . . . . . . . . . . . . 3450 6594
Exchange
1.
Open housing
2.
Shift type plate.
a)
Warm up type plate with a hair dryer until the adhesive
can be removed (not too hot as otherwise the housing is
damaged).
b)
Clean adhesive position on new housing and stick type
plate. New type plates can only be ordered as spare parts
if the old type plates are returned to B.Braun
ring of Spare Parts and Test Equipment“
3.
Remove drive from old housing
4.
Install drive in new housing
5.
Close housing.
Note
Do not squeeze the cable
6.
Calibrate in Service Program
Содержание 0871 4827
Страница 1: ...Service Manual Perfusor compact Version 2 1 english 0 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes ...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes ...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes ...
Страница 54: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb For your notes ...
Страница 60: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb For your notes ...
Страница 62: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes ...
Страница 69: ...Procedural Instructions on the TSC 8 8 5 Perfusor compact 2 1 gb ...
Страница 70: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb For your notes ...
Страница 72: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes ...
Страница 76: ...10 4 Perfusor compact 2 1 gb For your notes ...
Страница 80: ...Appendix A A A 2 Perfusor compact 2 1 gb For your notes ...
Страница 82: ...Modification Instructions for Syringe Holder B B 2 Perfusor compact 2 1 gb For your notes ...