![B+B Sensors 1M Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/b-b-sensors/1m/1m_operating-manual_460614071.webp)
OPEARTING MANUAL
Technische Änderungen vorbehalten
B+B Thermo-Technik GmbH | Heinrich-Hertz-Str. 4 | D-78166 Donaueschingen
0141 0315-100 24.06.2022
Fon +49 771 83160 | Fax +49 771 831650 | [email protected] | bb-sensors.com
71 / 88
4M Modell
Anzeige
Modus [Beispiel]
Einstellbereich
TPROC 320.9
Process temperature (after signal processing) [320,9 °C]
fixed
T INT 50.1
Detector Temperature [50,1 °C]
fixed
T BOX 38.6
Electronic box Temperature [38,6 °C]
fixed
EMISS 1.000
Emissivity [1,000]
0,100 ... 1,100
TRANS 1.000
Transmissivity [1,000]
0,100 ... 1,100
AVG 0.020
Signal output Average [0,020 s]
AVG 0.000 = inactiv/ 0,1 … 65 s
HOLD
OFF
OFF/ PEAK/ VALL/ APEAK/ AVALL
H TIM
PEAK/ VALL
0…65 s (65 = infinite)
H TH
APEAK/ AVALL
Starting temperature…end temperature
H HY
APEAK/ AVALL
Hysteresis setting in °C/°F
U °C
Temperature unit [°C]
°C/ °F
M 01
Multidrop address [1] (only with RS485 interface)
RS422 mode
01…32
RS422 (Press Down button on M01)
BAUD 115.2K
Baud rate in kBaud [115]
115.2 / 921.6 kBaud
S ON
Laser Sighting
ON/ OFF