Annex A
The product is suitable for use in an specific electromagnetic environment.
The customer and/or the user of the product should assure that it is used in an electromagnetic environment as described below.
Emission Test
RF-emission CISPR 11
RF-emission CISPR 11
Compliance
Group 1
Class A
Harmonic emissions IEC 6100-3-2
Class A
Voltage fluctuations/
flicker emissions IEC 61000-3-3
Complies
Electromagnetic Environment Guidance
The product use RF energy only for its internal function.
Therefore, its RF emissions are very low and not likely to cause any interference in nearby electronic equipment.
The product is suitable for use in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic purpose.
10.2 Electromagnetic Immunity
10.1 Electromagnetic Compatibility
The product is suitable for use in a specific electromagnetic environment.
The customer and/or the user of the product should assure that it is used in an electromagnetic environment as described below.
Emission Test
Electrical fast transient/bursts
IEC61000-4-4
Electrostatic discharge(ESD)
IEC61000-4-2
IEC 60601- Level
Compliance Level
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Surge
IEC61000-4-5
Electromagnetic Environment Guidance
Floor should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative humidity should be at least 30 %
Mains power quality should be that of a typical commercial and/or hospital environment
Mains power quality should be that of a typical commercial and/or hospital environment
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC61000-4-11
<5% UT (>95% dip in UT)
for 0.5 cycle 40% UT
(60% dip in UT)
for 5 cycles 70% UT
(30% DIP IN UT)
for 25 cycles <5% UT
(>95% dip in UT) for 5 sec
<5% UT (>95% dip in UT)
for 0.5 cycle 40% UT
(60% dip in UT)
for 5 cycles 70% UT
(30% DIP IN UT)
for 25 cycles <5% UT
(>95% dip in UT) for 5 sec
3A/m
3A/m
Mains power quality should be that of a typical commercial and/or hospital environment.
If the user of the product requires continued operation during power mains interruptions,
it is recommended that the product be powered from an uninterruptible power supply or
a batter
Power frequency magnetic fields should be at levels characteristic of a typical location in
a typical commercial or hospital environment.
91