5.8 Change in torque value
5.9 Change in speed
Torque Up/Down
Button
SPEED Up/Down
Button
Torque section on the display flashes upon pressing torque button.
Torque is controlled using the button for adjusting set value.
Torque setting mode is exited by pressing Torque button again or another button for another function or when the motor is operated.
Speed section on the display flashes upon pressing Speed button.
Speed is controlled using the button for adjusting set value.
Speed setting mode is exited by pressing Speed button again or another button for another function or when the motor is operated.
Change the torque value of motor, 5 N.cm increase.
Gear ratio Torque(Ncm) Gear ratio Torque(Ncm)
1:1 - 1:2 -
16:1 5~60 20:1 5~70
27:1 5~80 32:1 5~80
Gear ratio Speed(rpm) Gear ratio Speed(rpm)
1:1 200~40,000 1:2 400~80,000
16:1 12~2,500 20:1 10~2,000
27:1 7~1481 32:1 6~1250
5.10 Operation of implant engine
5.11 Auto Calibration
SPEED Up/Down
Button
Implant engine operates upon completion of all settings.
When motor operates by pressing foot switch, the letter 'R' on display and the surrounding border flickers in turn.
The torque and speed indicate the current figures and a beep sounds when altered torque value reach 90% of set value.
Motor stops when it reach 100% of set value.
Auto Calibration is performed upon simultaneously pressing both buttons longer than 2 seconds.
When users press the speed up and down button simultaneously for 2 seconds, auto calibration is executed.
When auto calibration is executed, the motor starts with a beep sound.
Auto calibration is completed when speed increases from 0 to 40,000 RPM on the display and the motor will stop when it's done.
Auto calibration is recommended when actual speed of the motor is different from speed on the display.
Auto calibration is recommended after sterilization.