Chapter 6. Maintenance
6.1 Manual cleaning
1) Separate the motor and Foot Switch connected in the Control Unit.
2) Prepare a cloth (preferably cotton) or soft brush moistened with isopropyl alcohol.
3) Clean foreign substance on the entire surface and in the gaps with cloth or brush soaked in isopropyl alcohol for at least 3 minutes.
4) Repeat the cleaning process if foreign substance is found.
6.2 Inspection
1) Inspect whether there is any foreign substance visible to the naked eye.
2) Check the operating condition.
3) Repeat the cleaning process if foreign substance is found.
4) Repeat the above procedure for every single use.
6.3 Sterilization
Sterilization Applicable Products : Motor, Motor connection cable
1) Disconnect the motor cable from controller.
2) Attach the motor cap to the handpiece connection of the motor.
3) Sterilize the motor equipped with the cap and motor cable either before or after use under the following conditions.
※
Sterilization by moist heat(134~135
℃
) for 3 minutes in a steam sterilizer (Autoclave)
Clean the product before and after use.
• Please sterilize after using the product.
• We do not recommend you to sterilize on plasma sterilization or
EOG sterilization.
• Do not disconnect the motor from the motor cable when sterilizing.
6.4 Change fuse
①
Switch off and unplug the unit as per 5.1
②
Carefully remove the fuse box
③
Replace the fuses with new ones
of the same fuse type.
④
Re-insert the fuse box
Extra fuse
Needed
to change
Type
Gravity displacement type sterilizer
Air-removal steam sterilization(pre-vacuum)
Sterilization conditions
At least 30 minutes at 121
℃
At least 4 minutes at 132
℃
Drying time
30 minutes.
30 minutes.
13