Operating Instructions BA ULTIMATE
7 Conditioning
65 56 083 D3560
.
201.10.02.02
29
Using tbe BA Ultimate spray
NOTICE! Using sprays from other manufacturers can reduce the
product's service life. Only use BA Ultimate Spray.
Use soft brush
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Procedure _ Turbine _ without Alpro
✔ Wear appropriate protective clothing.
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
1. Brush the turbine under running water (< 38 °C, < 100 °F, at least
drinking water quality) and good lighting (min. 500 lux) and color
rendering index (min. 80 Ra) until no more dirt can be seen, for at
least 10 seconds.
2. Flush the drive channels with spray.
3. Clean and disinfect spray channels with suitable agents and
adapters according to manufacturer's instructions.
4. Blow spray channels with suitable adapters out with 2.5 - 3 bar until
no more moisture displaces, but at least 10 seconds.
6556083.book Seite 29 Montag, 18. April 2016 4:35 16
Содержание BA 695
Страница 2: ...6556083_Titel fm Seite 2 Montag 18 April 2016 5 16 17 ...
Страница 4: ...6556083 book Seite 2 Montag 18 April 2016 4 35 16 ...
Страница 40: ...6556083 book Seite 38 Montag 18 April 2016 4 35 16 ...
Страница 42: ...6556083 book Seite 2 Montag 18 April 2016 4 30 16 ...
Страница 78: ...6556083 book Seite 38 Montag 18 April 2016 4 30 16 ...
Страница 80: ...6556083_FR book Seite 2 Montag 18 April 2016 4 40 16 ...
Страница 116: ...6556083_FR book Seite 38 Montag 18 April 2016 4 40 16 ...
Страница 118: ...6556083_ES book Seite 2 Montag 18 April 2016 4 51 16 ...
Страница 156: ...6556083_IT book Seite 2 Montag 18 April 2016 5 02 17 ...
Страница 192: ...6556083_IT book Seite 38 Montag 18 April 2016 5 02 17 ...
Страница 194: ...6556083_NL book Seite 2 Montag 18 April 2016 5 12 17 ...
Страница 230: ...6556083_NL book Seite 38 Montag 18 April 2016 5 12 17 ...