Azure™ cSeries User Manual

Page 13

  5. Select 

CAPTURE

. The channels will be captured sequentially. 

 

6.  The resulting image will be displayed in the gallery tab.

3.6 Custom Applications

The custom tab allows the users full flexibility when acquiring 

their image. 

        

 LOAD  PROTOCOL

 – Load Previously saved 

protocols, for quick image capture

          

SAVE PROTOCOL

 – If you have a custom 

setting you plan to use multiple times, save 

it, for quick recall in the future. 

Imaging workflow:
 

1.   Select the sensitivity/resolution for your setting. To use the full camera resolution, select 

LOWEST

.

 

2.  Select the light sources you wish to use. 

 

  a.  If you wish to take a single channel image, select a light source next to 

GRAYSCALE

 

  b.   If you wish to take a multichannel image, make sure the imaging time next to Grayscale is 

set to 

0

. Then, select the light sources you would like to image, next to the color channel 

you would like them assigned to. Set the exposure time for each channel. 

 

3.   Select the aperture setting: the smaller the number, the wider the aperture. For 

chemiluminescence, we recommend 0.95 for fastest light gathering. For all other applications, 

2.8 is recommend for improved depth of field and focus. 

 

4.  Select the focus setting:

 

  a.  For chemi applications on the Blot Tray on the UV transilluminator, we suggest 2600. 

 

  b.  For chemi applications on the Blot Tray on the chemi shelf, we suggest 8400.

 

  c.  For NIR applications, we suggest 3700-4200.

 

  d.  For RGB applications, we suggest 2400-3000.

 

  e.  For gel applications, we suggest 2400.

 

5.  Select the number of frames.

 

6.  Save your protocol if you think you will use it in the future. 

  7. Hit 

CAPTURE

. The resulting imaging will be displayed in the gallery tab.

3.7 Save Acquired Image

Users are discouraged to store image files on the tablet since 

its disk space and memory resources are limited. Please use 

a USB storage device, or if the system is networked, save to 

a network location. There are three USB port, one in the front 

and two on the right side panel.

Содержание c150, c200, c300, c400, c500, c600

Страница 1: ...Azure cSeries User Manual Page 1 Azure cSeries User Manual for the c150 c200 c300 c400 c500 and c600 NIR RGB CHEMI WHITE LIGHT BLUE LIGHT UV Rev 20150428...

Страница 2: ...coming in through the sides or around the lenses Always cover all skin that may be exposed to UV light especially the face neck hands and arms Always make sure that any UV protection devices such as t...

Страница 3: ...al illumination sources Right and left laser modules located in the c500 and c600 imaging chamber are installed in a highly diffused protective metal enclosure preventing exposure to full laser radiat...

Страница 4: ...however are specifically excluded Defects caused by improper operation Repair or modification done by anyone other than Azure Biosystems or the company s authorized agent Use of spare parts supplied b...

Страница 5: ...3 2A Imaging Workflow 9 3 2B Band Excision and Light Safety Switch 10 3 3 Chemiluminescent Applications c300 systems and above 10 3 3A Imaging Workflow for Chemiluminescent Applications 11 3 4 RGB App...

Страница 6: ...5nm trans illumination It is mounted on a pull out tray EPI Blue and White LEDs The cabinet also contains EPI white and blue light panels on each side of the cabinet interior walls EPI RGB LEDs High i...

Страница 7: ...ter solvents or corrosive materials on a flat and stable surface with adequate clearance on all sides The system must remain stationary during operation Further the system should be placed away from i...

Страница 8: ...ng screen during initialization Once the system is done initializing the software displays the gallery tab and is ready to use 2 5 Attach Other USB Input Devices to the System You may attach regulator...

Страница 9: ...ll display your image with dark and light pixels reversed SATURATION Selecting Saturation will show you images with intensities beyond the dynamic range of the camera SINGLE MULTICHANNELVIEWING When a...

Страница 10: ...ht source for gels that are stained with dyes that excite in the blue range Visible For dyes that are visible without additional excitation use the Visible light source When selected the software will...

Страница 11: ...he resolution of the camera and image At the Lowest sensitivity setting you will be taking a full resolution image with the camera And the Highest sensitivity setting the resulting image will be 0 3 M...

Страница 12: ...amera Recommendation Set the exposure time to 30 Seconds by hitting 3 then 0 and then S Keep the Sensitivity setting on the default setting NORMAL Select CUMULATIVE After hitting CAPTURE a series of i...

Страница 13: ...e time 5 Select CAPTURE The channels will be captured sequentially 6 The resulting image will be displayed in the gallery tab 3 5 NIR Applications c500 and c600 For all NIR blots use the Blot Tray on...

Страница 14: ...d like them assigned to Set the exposure time for each channel 3 Select the aperture setting the smaller the number the wider the aperture For chemiluminescence we recommend 0 95 for fastest light gat...

Страница 15: ...CUS CALIBRATION Allows the technician to reset the focus setting in case of a shift in focus CREATE FLATS Allows the technician to recreate flat calibration files 5 Help Selecting the help tab will al...

Страница 16: ...ower Off Restart the Tablet To turn off the system shut down the tablet by a Swiping from the right to left on the right side of the screen to bring up the right side toolbar b Select SETTINGS c Selec...

Страница 17: ...st of programs to uninstall select CSERIES CAPTURE SOFTWARE 4 Once it is uninstalled you can begin installing the new software 5 Select the SETUP file in the new software folder 6 Select RUN from the...

Страница 18: ...the default 12 If it asks you to replace the existing cSeries folder select YES then INSTALL 13 On the CP210x USB to UART Bridge Driver Installer screen select NEXT 14 On the next screen accept the ag...

Страница 19: ...the software and make sure the hardware is turned on and re launch the software Copyright 2013 Azure Biosystems Inc All rights reserved The Azure Biosystems logo and Azure are trademarks of the Compan...

Отзывы: