While in calibration HI mode, the default HI value will be displayed
on LCD as 1.600 or 4.000... based on different models. While in the
calibration LOW mode, the default LOW value will be displayed as
well.
2. If your standard calibrator is not the same as meter default value,
press up/down or number keys to adjust. The adjustable range is
displayed on LCD as a reminder. Please refer to table in page 36
for the details. The ZERO calibration could not be adjusted and the
value always should be 0.000
3.After adjust the value to be complaint with calibrator, press F4 to
start the calibration.
4. While doing HI calibration, please connect positive end tube to
calibrator. While doing LOW calibration, please connect negative
end rube to calibrator. While doing ZERO calibration, no need to
connect tube to any calibrator.
5. After 5 seconds calibration, OK or NG will be displayed on the LCD
to indicate the calibration is successful or not. If OK, the actual
measured value which is displayed above the text "OK" Should be
within the specification.
6. If the NG displayed, try the calibration process again. If get three
time failure, the meter need to be returned for repair
7. While finishing one point calibration, press EXIT to back to main
calibration mode. Then, choosing to exit the calibration or entering
to next point calibration. Maximum three points calibration to have
accurate measurement.
37