EDIT
PRN
EXIT
NEXT
Set Clock:
MM-DD-YY HH:MM:SS
Set ID:
- - : :2 0
06 30 05 18 29
Disable
ID: User Name
EDIT
PRN
EXIT
NEXT
LCD Cont.(1-5):
Prn Cont. (1-9):
Unit:
Date:
Auto Off:
3
5
YY-MM-DD
20mins
Enable
Metric
EXIT
EDIT
BACK
PH
Volt
Temp
Select Item
PRN
MEAS
MEM
LOG
SET
Welcome
Volt:
EXIT
CAL
PRN
Temp:
PH:
28C
7.60
-34.9mV
ATC
EXIT
LXW
ENTER
SAVE
Volt: -3.6mV
PH: 7.00
"CALIBRATION"
Auto Judging...
Last: 76 days ago
RST
Remarks:
9861/9661 pH/mV Setting / Single measurement
1.No PRN function is existing for model 9661 meter.
2.Press CAL to enter calibration mode , "Auto judging...Last
xx days " means it has been xx days after the last calibra-
tion .
3.If you see a minus days "...- xx days " means you haven't
set the real time. Set real time from the main screen page.
4.Calibration is necessary and should be done regularly, re-
commend everyday if the meter is used often.
5.Meter features automatic buffer recognition to avoid errors.
24
Long press