9812/9612 External Temp. Probe Calibration procedure
MEAS
MEM
LOG
SET
Welcome
C
EXIT
PRN
Press ON to Ir Meas
Ext:30.5
Ir:
CALI
Long press
EXIT
LXW
EDIT
SAVE
Probe Cali Data
Step 1 :0 at 5C
Step 2: -5 at 40C
Step 3: -6 at 85C
Note: Unit is 0.1C
ABORT
LXW
ENTER
SAVE
Input Cali Data
Step 1 :+
00
+/-
ABORT
LXW
ENTER
SAVE
Input Cali Data
Step 2 :+
00
+/-
Repeat until three points
value are saved
Remarks:
1. Be sure to keep a calibration
note enclosed in the package
This's for future probe calibration
use
2. While input the calibration value, refer to the value indicated
on the note. There are three values for different temperature.
The value is unique for each probe.
3. To exit manual calibration mode, long press F1 anytime to
exit.
4. +/- of the calibration means positive value or negative value.
5. Use 10 number keys of the meter to input the calibration value.
19