28
OPERATION MANUAL
__________________________________________________________________________________________________________________
17. Maintenance and repairs of small defects
1. The scale should be kept clean.
2. Take care that no dirt gets between the platform and the scale base. If found
any, remove the pan (lift it up), remove dirt and then replace the pan.
3. In case of improper operation caused by short-lasting power supply decay,
unplug the scale from the mains and then plug it again after few seconds.
4. It is forbidden to make any repairs by unauthorised persons.
5. To repair the scale, please contact our nearest service. List of authorized
services is show on site: www.axis.pl.
6. Damages scales should be sent to repair only in original package. Scale should
be protected against pressure.
Error communicates:
Communicate
Possible cause
Remedy
C-1
...
6
(over 1 min.)
self test failed
if displayed more than 1 minute,
contact an authorised service
balance
doesn’t work
protection rod
remove protection rod and cap
L
pan missing
put the pan on
mechanical damage
contact an authorized service
H
overloading
remove the load from the pan
mechanical damage
contact an authorized service
the scale was switched on with
loaded pan
remove the load from the pan
unstable ground
vibrations
air flows
place the balance on a stable
ground not affected by
mechanical vibrations and
airflows
indicator does
not appear
balance damage
contact an authorized service
-
-
-
-
-
-
taring is progress
contact an authorized service