Operation rules during work with accumulators (batteries)
1.
Scale can be powered from ~230V supply through feeder attached with scale. Moreover accumulators,
which are placed in container inside the scale, can be used for powering. Ordinary batteries can be
used as well.
2.
In order to make accumulators (batteries) discharging time longer, automatic switching off display
backlight and next the whole scale is possible during breaks in weighing. Configuration of these
mechanisms is done using
b_LIGHT
and
Auto OFF
functions.
3.
Charging accumulators is performed automatically after connecting feeder to the scale, also during
weighing. Accumulator power level can be read using
bAttErY
function (
bat VoL
option).
11.
Scale checking
It is advised to check scale indication accuracy before and after series of measurement using any load with
known weight.
To check the scale with legal verification use a calibration weight with valid calibration certificate. In case
permissible error is exceeded it is advised to contact the nearest service to calibrate the scale.
When using batteries in place of accumulators, charging during work with feeder have
to be switched off.
bAttErY
function is used for this purpose (
bAt OFF
option), which is
described in further part of manual. Charging batteries can cause their breaking and
serious damage of the scale.