14.18
External calibration (CALIb)
Note:
This function is enabled in non legalized scales only.
Calibration of sensitivity should be make when aaccuracy of scale is not
satisfactory. Could use standard mass equal of maximum range of scale (
Max
).
Operation sequence:
Press
MENU
key to call functions
menu and to choose
CALIbr
with
→
T
←
key.
The
following
options
appear
successively on display:
-
CAL-0
– leave without calibration,
-
CAL-1
– quick calibration – without
confirms by
MENU
key,
-
CAL–2
– calibration with confirms
-
out
– leave without changes
Press
MENU
when communicate
CAL-2
is display.
Confirm readiness to calibration by
press
MENU
key – pan must be
empty.
When communicate C is display put
a standard mass (equal of maximum)
to a pan and press
MENU
key.
Wait to end of calibration process.
m
0.00
CAL
0.00
m
0.00
CALIb
CAL-2
CAL-0
...
T
T