USER MANUAL
49
____________________________________________________________________________________________________________________
Application options:
- Activation
–
move to density measurement,
-
Preview
–
preview of actual summing register,
-
Material type
–
solid body or liquid,
-
Plunger volume
–
insert volume marked on
plunger,
-
Shortcut
–
fast access key by navigation keys
select: F1, F2, ... or F5.
Sequence of actions for liquids:
After choosing liquid as the material type,
inscribing plunger volume and selecting
Activation
tare the scale with
T
key
.
Hang the plunger without dipping into liquid
(measurement in air) and press
ENTER
.
Hang the plunger and dip it into liquid
(measurement in liquid)
and press
ENTER
.
Results are displayed and following options:
- result printout,
CLR - exit to summing
.
In order to finish work with application, choose
application from menu and select Deactivation.
Max ... Min ... e= ... d= .. II
AUT DEN
T
0.000
g
0%
100%
Max ... Min ... e= ... d= .. II
AUT DEN
ENTER
10.090
g
0%
100%
Max ... Min ... e= ... d= .. II
AUT DEN
ENTER
8.091
g
0%
100%
DENSITY/PREVIEW
Mass in air
: 10.090g
Mass in liquid : 8.091g
Liquid density : 5.037g/cm3
Density with comp.. : 5.036g/cm3
- printout CLR - exit
DENSITY
1. Activation
2. Preview
3. Material type:
<Solid>
<Liquid>
4. Plunger volume: ......
5. Liquid density:
..... g/cm3
6. Shortcut:
<-><F1><F2>...<F5>
7. Exit