62
USER MANUAL
__________________________________________________________________________________________________________________
Scale
Computer: without response,
Scale turning on / off (calling
I/
key press):
Computer
Scale: S S CR LF (53h 53h 0Dh 0Ah),
Scale
Computer: without response,
Entering to special function menu (calling
MENU
key press):
Computer
Scale: S F CR LF (53h 46h 0Dh 0Ah),
Scale
Computer: without response,
Setting low threshold value (option):
Computer
Scale: S L
D1...DN
CR LF (53h 4Ch
D1...DN
0Dh 0Ah)
D1...DN
–
threshold value, maximum 8 characters
(„
-
” –
negative value, digits, dot
–
decimal
separator), number of digits after dot should be the same as on scale display,
Scale
Computer: without response,
Example:
in order to set low threshold 1000g in scale B1.5 (d=0.5g) the following order should be sent:
S L 1 0 0 0 . 0 CR LF (53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah),
in order to set low threshold 100kg in scale B150 (d=50g) the following order should be sent:
S L 1 0 0 . 0 0 CR LF (53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),),
Setting high threshold value (option):
Computer
Scale: S H
D1...DN
CR LF (53h 48h
D1...DN
0Dh 0Ah),
D1...DN
–
threshold value (see )
Scale
Computer: without response.
Connecting cable WK-1
(scale
–
computer / 9-pin interface):
Connecting cable WD-1
(connects printer with scale):
AXIS C-001 printer internal switches setting:
SW-1
SW-2
SW-3 SW-4 SW-5 SW-6 SW-7 SW-8
on
off
on
off
off
on
off
off
SCALE
COMPUTER
SCALE
PRINTER