26
USER MANUAL
____________________________________________________________________________________________________________________
After setting standard of mass value, prepare
standard of mass, choose
Calibration
option
and press
ENTER
key.
Put standard of mass on pan.
Standard of mass value indication
means that calibration process ended.
Besides
Report
option, all other calibration options are available after switch position
change.
SETUP
1. User
2. Menu
3. Calibration
4. Auto-zeroing
5. Unit
6. Interface
7. Print setup
8. LCD settings
9. Language
10. Time&date
11. Keyboard
12. Analog output
13. Firmware update
14. Defaults
14. Exit
CALIBRATION
1. Calibration:
2. Mode
3. Load: 100g
4. Report
5. Linearity
6. Exit
CALIBRATION
Please wait ...
CALIBRATION
Taring
CALIBRATION
Put load
CALIBRATION
Please wait ...
Max ... Min ... e= ... d= ..
100.0000
g
0%
100%