Appendix A
Information’s concerning double-range scale (options)
1. General description
Double-range scale’s have capability of work with greatest accuracy in bottom
measuring range part. Weighing of smallest mass is more precise.
This type of scale’s have two measurement range:
-
Max
1
- 50 % of maximum load (mostly),
-
Max
2
– 100% of maximum load,
and adequate reads digit: d
1
i d
2
(d
1
< d
2
).
Double-range options causes change of scale’s operation:
-
after turn on (in small mass range 0- Max
1
) scale displays result with reading
unit d
1
,
-
when the load pass the Max
1
scale changes reading unit on d
2
; from this
moment scale displays result with reading unit d2 on all measure range
(0- Max
2
),
-
return to unit d
1
is succeed after zeroing the scale (
→
0
←
key), or when all
mass is removed from pan (indicator "
→
0
←
").
2. Double-range scale parameters
Range and graduation values are represented on nominal table on the scale.