CONSTANT CURRENT MODE 

1.

 

Turn the voltage regulator clockwise to maximum position and the current regulator anti – 

clockwise to minimum position. 

2.

 

Press the power ON/ OFF switch to ON.

 

3.

 

Connect the positive output terminal and negative output terminal with a load or similar component.

 

4.

 

Turn the current regulator clockwise to that you are desirous of output current value.

 

5.

 

The voltage regulator controls the output voltage indicator. The current regulator controls the output 

current indicator. Indicator show that output voltage and current of the output terminals.

 

 

RESTRICTED CURRENT PROTECTION MODE

 

1.

 

Press the power ON/ OFF switch to ON. 

2.

 

Turn the current regulator anti-clockwise to minimum position then clockwise a little. 

3.

 

Turn the voltage regulator clockwise to that you are desirous of output voltage level position 

(approx.1.5V). 

4.

 

Link the positive and negative output terminals with a wire. 

5.

 

Turn the current regulator clockwise to get the current level at which the restricted current 

protection will be active as you desire. 

6.

 

Remove the wire linked to the positive and negative output terminals. Connect the positive 

output terminal and negative output terminal with a load or similar component. 

7.

 

When the output current achieves the setting value of restricted current protection, and the 

restricted current protection is active. 

NOTE: 

After setting of the restricted current protection, if the current regulator is turned again, 

the restricted current protection range will be changed. 

C.C. AND C.V. INDICATOR 

(Only for the instrument with C.C. and C.V. INDICATOR )

 

The C.C. indicator is controlled by the constant current mode. Otherwise C.V. indicator is 
controlled by the constant voltage mode. 

CAUTION 

The DC power supply series has perfect restricted current protection, Even so, when the output 
terminal is short - circuit, the DC power supply should be turned off and the short - circuit should be 
removed before continuing operation because the power transistors in the instrument will bear 
heavy. The AC power must be switched off before servicing and servicing should be referred to  
a qualified person. 

 

Содержание AX-1803D

Страница 1: ...DC POWER SUPPLY AX 3003D AX 3005D AX 1803D USERS MANUAL V C V I N E C O AR SE VO LTAG E A V OFF _ ON VOLTAGE CURRENT DC POWER SUPPLY F I N C O AR SE F I N E C R R EN T C C O ER O O N...

Страница 2: ...recautions 4 Safety Information 4 Safety Symbols 4 Instrument Front Panel 5 Instrument Back Panel 6 2 Operation 7 Line Power Supply Setting 8 Operation Procedure 8 3 Specifications 10 General Specific...

Страница 3: ...ypes for the power supply series monitoring output voltage and current They are LCD LED and two pointer meters One of them is equipped to the DC power supply There are different numbers of output volt...

Страница 4: ...taking electrical measurements Keep your body isolated from ground by using dry clothing rubber shoes rubber mat or any approved insulating material 5 Never touch exposed wiring connections or any liv...

Страница 5: ...1 1 OUTPUT VOLTAGE INDICATOR This indicator shows output voltage as measured at the output terminals There is LCD or LED or pointer meter display type They are available for choice of user 2 OUTPUT C...

Страница 6: ...only pointer meter display Each pointer meter has a mechanical screw adjustment for setting the zero point Using a small screwdriver turn off the power and adjust the screw under the meter respective...

Страница 7: ...with same the power requirements of instrument before connect power cord to live power source WARNING To avoid electrical shock disconnect power cord from live power source and remove the test leads...

Страница 8: ...50Hz 60Hz NOTE The instrument without POWER SELECT SWITCH is factory pre set in accordance with your main voltage when your local distributor place the order with the factory Operating Procedure CONST...

Страница 9: ...rrent level at which the restricted current protection will be active as you desire 6 Remove the wire linked to the positive and negative output terminals Connect the positive output terminal and nega...

Страница 10: ...ous upward of 75 rate output power for one hour Relative Humidity 20 to 80 RH 0 to 40 Protection constant current and short circuit protection Weight Size OUTPUT VOLTAGE regulated OUTPUT CURRENT regul...

Страница 11: ...0 2A 0 3A 0 18V 0 2A 0 3A 0 30V 0 2A 0 3A 0 5A 0 10A 0 20A 0 50V 0 2A 0 3A Source Effect Line regulation CV 0 01 1mV CC 0 2 1mA Loading Effect Line regulation CV 0 01 3mV CC 0 2 3mA Ripple nad noise C...

Страница 12: ...TO V CAUTION WARNING Turn off the power switch remove the power line cord from the power socket and disconnect the test leads at output terminals before replacing the fuse Replace it only with same ty...

Отзывы: